Hydrogen peroxide (H2O2) is a key molecule in numerous physiological, industrial, and environmental processes. H2O2 is monitored using various methods like colorimetry, luminescence, fluorescence, and electrochemical methods. Here, we aim to provide a comprehensive review of solid state sensors to monitor H2O2. The review covers three categories of sensors: chemiresistive, conductometric, and field effect transistors. A brief description of the sensing mechanisms of these sensors has been provided. All three sensor types are evaluated based on the sensing parameters like sensitivity, limit of detection, measuring range and response time. We highlight those sensors which have advanced the field by using innovative materials or sensor fabrication techniques. Finally, we discuss the limitations of current solid state sensors and the future directions for research and development in this exciting area.
H2O2 plays an important role in various applications such as medical diagnostics, clinical research, and industrial sectors like food processing, paper, textile, pharmaceuticals as well as cleaning and disinfection products (Figure 1) [1]. H2O2 is also important physiologically and is involved in metabolic activities, apoptosis, and immune cell activation [2][3]. It plays an important role as an oxidative stress marker, defense agent, and aging [2][4]. It is a crucial biomarker in monitoring various diseases and disorders including diabetes [5], cancer [6], Parkinson's [7], cardiovascular, Alzheimer's [7], and neurodegenerative disorders [7][8]. Moreover, H2O2 is the intermediate molecule formed in reactions involving numerous oxidases such as glucose oxidase, alcohol oxidase, cholesterol oxidase, lactate oxidase, and glutamate oxidase [9]. Further, H2O2 is used for sterilizing various medical equipment and residual H2O2 levels need to be monitored to ensure that the equipment is safe to use [10].
Figure 1. Overview of H2O2 detection with inner circle containing the two common detection principles: optical and electrochemical; and outer circle with few applications areas of H2O2 detection.
H2O2 measurement and quantification is performed in a variety of sample matrices including environmental samples like water and soil, human fluids like sweat, blood, cell and tissue cultures. H2O2 is measured using diverse range of methods such as optical [11][12] including colorimetry, chemiluminescence, and fluorescence; and electrochemical [13][14][15][16] including potentiometry, voltammetry and amperometry (Figure 1). Optical techniques are limited by high cost, complex testing processes, the requirement of sophisticated and bulky instrumentation, need for trained personnel to operate, and interference from sample matrices. On the other hand, electrochemical sensors offer low-cost, simple instrumentation and fast detection [13]. Nevertheless, electrochemical sensors also suffer from a few limitations such as the requirement for a reference electrode, larger working area, etc. The potentiometric method requires a reference electrode for reliable potential measurement while amperometric sensors require a reference electrode to apply a reliable potential bias for the measurement. For potentiometric sensors, a stable response strongly depends on the stability of the reference electrode. However, a miniaturized solid-state reference electrode with long term stability is yet to be realized [17]. For amperometric sensors, a high working electrode potential results in increased interference from interfering molecules [18].
More recently, solid-state sensors such as chemiresistors [19][20][21], conductometric sensors [21][22][23] and field effect transistors (FET) [24][25][26] have been used to measure H2O2 while avoiding the aforementioned challenges. Chemiresistors consist of a single sensing layer which measures the change in analyte concentration through alteration in resistance of the layer using two contact electrodes. A small potential bias is applied to the substrate film and the change in current is measured. Advantages of chemiresistors are: high sensitivity, because the resistance changes can occur due to modification at any position of the network unlike techniques like colorimetric which is based on volume modifications; ease of fabrication of sensor arrays due to simple sensor structure; suitability for miniaturization; simple instrumentation setup for measurement and elimination of the need for reference electrodes unlike electrochemical methods [27]. FET based solid state sensors are attractive due to their ability to detect analytes with ultrahigh sensitivity. In addition, FETs can be manufactured easily using the established manufacturing process for metal oxide semiconductor FETs (MOSFET) [9].
Previous reviews on sensors for H2O2 detection have typically focused on electrochemical and colorimetric sensors. Several papers have been published on enzymatic [1][13][14][16][28][29] and non-enzymatic sensing [1][15][30][31] using those principles and the readers are referred to them for an in-depth analysis in these areas. An in-depth review of the emerging class of solid state H2O2 sensors is not currently available. This review is focused exclusively on chemiresistive, conductometric and FET based H2O2 sensors which have significant potential for field deployment. A critical analysis of the sensing methods with emphasis on the sensing mechanisms and important parameters like measuring range, limit of detection (LOD), and response time have been provided. The diverse range of functional materials used for sensing and to fabricate these sensors have also been discussed.
Chemiresistive sensors are a group of sensors which transduces the chemical changes to resistance change. The sensor response is attributed to surface reactions or adsorption of analyte molecules on the sensing film [32]. This type of sensor was originally developed for gas sensing by monitoring resistance changes with adsorption of gas molecules on the sensor surfaces [32][33]. Typically, a sensor is placed under a small potential bias and the change in current is measured as output and converted into a change in resistance. A general chemiresistor consists of four components: the sensitive or active thin film substrate, contact electrodes, passivation layer and substrate (Figure 2a,b). Although for gas sensing, the contacts may be exposed to the environment, they are typically covered with an insulating film to avoid electrical shorting, especially when used in conducting liquids. The equivalent electrical circuit for a chemiresistive sensor can be represented as shown in Figure 2c, where both contacts are represented by parallel RC circuit depicting both Faradaic and non-Faradaic processes. The sensing layer which remains in contact with the solution is divided into three parallel RC circuits representing surface, bulk, and interface processes. When chemiresistive sensors are operated in DC mode, all capacitance can be neglected from the equivalent circuit (Figure 2c).
Figure 2. (a) Top view of a chemiresistive sensor with 4 main components: active material (dark grey), connectors or contacts (gold), dielectric to insulate the contacts (grey) and substrate (light grey) (b) A transverse section of a chemiresistive sensor with two connecting outputs (c) An electrical circuit analog for the chemiresistive sensor with Rc1 and Rc2, the contact resistance for the first and second contacts, while RS, RB and RI are the solution, bulk and interfacial resistance. Similar to resistance, capacitance of all the surfaces are labelled accordingly.
During measurement, the sensor is exposed to the analyte, and adsorption of analyte to the active thin film results in a change in resistance. For instance, carbon nanotubes (CNTs) are generally p-doped when the films are coated using water based CNT dispersions and if analyte adsorption results in the release of electrons, the hole concentration in the active surface is reduced, which results in a decrease in resistance [34]. On the other hand, if the analyte extracts electrons from the CNT film, this will lead to an increase in dominant carriers resulting in an increase in conductivity. Further, this change in resistance due to analyte interactions can occur from factors like increasing the CNT-CNT junction resistance modulation of the Schottky barrier at the CNT-metal contact junction and charge transfer between analyte and CNT. These processes have been described in detail in other reviews [35][36]. These sensors have some limitations such as irreversible changes introduced onto the substrate due to application of a potential bias, a high dependence of the sensitivity of the sensor on the substrate thickness, and high contact resistance which can further reduce the sensitivity of the sensor. For instance, in the case when conducting polymers are used as the functional sensing layer, the potential bias can induce an irreversible change in the polymer film resulting in a change to the baseline resistance of the sensor. The analyte can also cause irreversible changes to the sensors surface [19]. Thinner films generally have higher sensitivity as compared to thicker films [33]. For two point measurements, the resistance change has two components: change due to the analyte binding and change in contact resistance between the substrate and the metal contacts [37].
Conductometric sensors are devices which detect the change in conductivity of the analyte solution due to consumption or generation of ions due to chemical reactions using two conducting electrodes [38]. This method was originally developed to study chemical kinetics of reactions and later exploited by researchers to detect enzyme catalyzed reactions. Conductometric measurements are non-specific as conductivity changes can occur due to the migration of all ions present in the solution. This non-specificity is circumvented by coating enzyme on top of the electrode and doing the measurements in a defined measuring cell. Conventionally, the conductivity measurements are performed in AC mode. Unlike chemiresistive sensors, these sensors offer information through the frequency of the measurement, an important experimental variable to determine non-Faradaic processes. An alternating bias has several advantages such as minimized contact polarization, double layer charging and electrode polarization [39].
Typically, conductometric measurements are done using a pair of identical electrodes (generally interdigitated electrodes) dipped in a solution container with a constant volume. One of the interdigitated electrodes (IDE-1) is coated with the enzyme film and the other does not have any enzyme layer (IDE-2) (Figure 3a). The IDE-2 determines the base conductivity response from other ions and molecules present in the solution. The measurement of both the sensors are done with respect to a counter and/or reference electrode (Figure 3b). The final sensor response is determined by subtracting the signal of IDE-1 from the signal of IDE-2. Here, the impedance is measured perpendicular to the electrode surface. The equivalent circuit of the electrochemical cell is shown in Figure 3c where Rct1 and Rct2 are the charge transfer resistances for IDE and CE respectively, W is the Warburg impedance for the IDE which models the diffusional resistance due to both interfaces, Cdl1 and Cdl2 are double layer capacitances of IDE and CE respectively and Rs is the solution resistance. Enzymatic conductometric sensors are versatile sensors which are low-cost, need a smaller potential bias and require simple instrumentation to generate reliable signals. However, the sensing signal can be affected by temperature variations [39] and changes in the ionic strength of the solution.
Figure 3. (a) A pair of interdigitated electrodes (IDE): IDE-1 represents the active membrane coated electrode and IDE-2 is the electrode coated with membrane without any active material (b) A schematic representing the experimental setup of a conductometric sensor with IDE, a counter electrode and a reference electrode. The zoomed in picture shows a transverse section of IDE and counter electrode (c) An equivalent circuit for the conductometric sensor with RC1, W and Cdl1 representing the charge transfer resistance, Warburg impedance and double layer capacitance, respectively for the IDE; RS is the solution resistance; and RC2 and Cdl2 representing the charge transfer resistance and double layer capacitance for the counter electrode respectively.
Metal oxide field effect transistors (MOSFETs) are used in electronic circuits as switches, gates, amplifiers etc. MOSFETs can be three or four terminals depending on the presence or absence of back gate (base substrate): source, drain, gate and base substrate. Insulated gate FET (IGFET) is the most common type of MOSFET used currently for chemical sensing. The gate terminal of the IGFET is insulated using a dielectric layer (like SiO2). A typical n-channel FET is constructed using a p-type substrate with heavily doped n-type source and drain (Figure 4a). The operation of the FET depends on the potential bias applied to the gate. Under zero bias, the FET channel is non-conducting. For n-channel FET, the conduction begins after a critical threshold potential is applied to the gate. This threshold potential will induce an inversion layer.
Figure 4. Schematics of a (a) MOSFET with p-type silicon as a base substrate (pink) and n-doped source and drain region (green) (b) ISFET with a pH sensitive film (orange) (c) Back gated FET with analyte solution and RE on the top side and all the terminal connections are done from the back side (d) Extended gate FET with a regular MOSFET and an extended sensing region connected with gate terminal of the MOSFET. S, G, D and B are source, drain, gate and base substrate terminals (all shown in black). RE is the reference electrode.
Early H2O2 (and glucose) FET sensors had pH sensitive material coated on the gate insulator that made it sensitive to changes in local pH due to generation or consumption of hydrogen ions by an enzyme that catalyzes H2O2 (Reaction 1) [40][41][42]. In this reaction, the reduction of H2O2 was catalyzed in presence of horseradish peroxidase (HRP), with the iodide ion acting as a reducing agent [43].
In such FETs, the gate dielectric is converted into a hydrogen sensitive film which can generate similar potential change in presence of the analyte (Figure 4b). Then the channel conduction can be influenced by changes in the hydrogen ion concentration. These devices are known as ion selective FETs (ISFETs). Similar to MOSFETs, ISFETs can also be n-channel or p-channel ISFET depending on the doping of the silicon substrate used to fabricate the FET. The drain current depends on the resistance of inversion layer and, the potential applied between source and drain. Mathematically, the drain current (Id) of ISFET is given by [9]:
where μ is mobility of electrons in the channel; L and W are length and width of the channel, respectively; q is the elementary charge, Qb, Qi, Qss are the charges located in depletion region, insulator region, and surface and interface states, respectively, χsol is solution's surface dipole potential, Eref is the reference electrode's potential, ϕSi is the electron work function of silicon, ϕf is the potential difference between Fermi level of doped and intrinsic silicon, Vds is the potential applied to the drain with respect to source, Vd is the drain potential, Vg is gate potential, Ci is the capacitance value of the gate, and ϕ is the potential of membrane–electrolyte interface.
Commonly used pH sensitive materials for ISFETs are silicon nitride (Si3N4), aluminum oxide (Al2O3), and tantalum oxide (Ta2O5) [9][40][41][42][43][44][45][46]. pH sensitivity is a surface phenomenon where the surface hydroxyl groups interact with the protons. The consequent changes in the surface charge or potential of the gate material leads to a current flow in the channel. This generated current is proportional to the analyte concentration. ISFETs are increasingly popular due to advantages such as rapid detection, small size, established manufacturing process, easy integrability in arrays, and their ability to be stored in dry form. However, they suffer from higher drift as compared to ion selective electrodes.
Based on its design, the ISFETs can be front side and back side connected. Front side ISFETs are widely used due to ease of fabrication, but they make it difficult to passivate the device from the analyte solution. Back side ISFETs solve the passivation problem as all the connections are accessed from the back side of the silicon chip. However, it poses a manufacturing difficulty to etch a deeper cavity into the silicon chip for connecting the source and drain from the back side of the chip (Figure 4c). An alternate configuration known as an extended gate FET was proposed in 1983 [47] (Figure 4d). The device has two components: a MOSFET with electrical connections and an extended gate with a pH sensitive film. This device has advantages such as low manufacturing cost due to simpler fabrication and packaging, and long term environmental stability of the device as the FET is not directly exposed to the solution.
Chemiresistive sensors were initially developed to detect gases or vapors but in the past two decades several chemiresistive sensors have also been developed to measure analytes in liquid environments. These sensors consist of two main components: an active thin film and electrical contacts. One of the first H2O2 chemiresistive sensor was fabricated from polypyrrole and multiwalled CNT (MWCNT) [48]. Chemiresistive sensors can be broadly classified based on active sensing thin film material. The active thin film can be made of various conducting or semiconducting materials such as CNTs [19][48][49], conducting polymers [20][50][51] or combinations of these materials. Contact electrodes are made of conductive materials including conductive carbon [50][51], metal electrodes like platinum [48], gold [19][20][52], and silver [49]. The sensors are compared based on three parameters: measuring range, LOD and response time. A summary of H2O2 chemiresistive sensors is given in Table 1.
Table 1. A list of H2O2 chemiresistive sensors with the crucial sensor properties including LOD, measuring range, voltage bias, response time, buffer and working pH. All units are as mentioned in the top row unless specified. Where NR: Not reported, PPy: Polypyrrole, PANI: Polyaniline, Pt NPs: Platinum nanoparticles, SnO2: Tin oxide, Au: Gold.
Substrate |
Target analyte |
Ligand/ Enzyme |
LOD (mM) |
Measuring range (mM) |
Voltage bias (mV) |
Response time (s) |
Buffer/working pH |
Comments |
Interference tested |
Ref |
Carbon nanotube based |
||||||||||
PPy-MWCNT |
H2O2/ Glucose |
Dodecyl benzene sulfonate |
NR |
0–20 |
1 |
NR |
NR |
Investigated the sensitivity of temperature humidity etc. |
No |
[48] |
CNT |
Glucose |
EGCG-GOD |
8.7 nM |
10 nM–1 μM |
100 |
<400 (est.) |
Working pH 7.4 Buffer: PBS |
Sensor responds to all reactive oxidative species |
Yes |
[49] |
SWCNT-PVP |
Glucose |
GOD |
0.08 |
0.02–2 |
100 |
3 |
Working pH 5.5 Buffer: Acetate |
Tested in juice & iced tea Stable for 5 consecutive tests |
Yes |
[19] |
Conducting polymer based |
||||||||||
Au-PANI nanowires |
H2O2 |
AgNPs |
5 |
5–40 |
20 |
25 |
Working pH 5 Buffer: Phosphate (200 mM) |
Stable response for 36 h Reusable sensor |
Yes |
[20] |
MWCNT-PANI nanowires |
H2O2 |
AgNPs |
1 |
1–20 |
NR |
180 |
NR |
Inkjet printed sensors |
No |
[51] |
MWCNT-PANI nanowire |
H2O2 Glucose |
PtNPs |
2 |
2–10 |
500 |
240 |
NR |
Inkjet printed |
No |
[50] |
Others |
||||||||||
Alumina |
Glucose |
SnO2-GOD |
0.5 (est.) |
0.5–20 |
NR |
50 |
Working pH 7.2 Buffer: Phosphate |
Sensor sensitivity increases with deposition temperature of SnO2 |
No |
[53] |
One of the earliest conductometric sensors for H2O2 was developed in 1999 [43]. The sensor was constructed using tetra-tert-butyl copper phthalocyanine (ttb-CuPc) coated gold interdigitated electrodes on a ceramic substrate (Figure 5a). The phthalocyanine film was deposited using the Langmuir-Blodgett method. The sensing of H2O2 was based on oxidation of iodide ions by H2O2 in the presence of HRP (Reaction 1). The iodine concentration was measured by a conductometric sensor based on ttb-CuPc as the sensing layer. The effect of other interfering species in the sample was suppressed using a hydrophobic gas permeable membrane. HRP was immobilized on top of the gas permeable membrane. The sensor attained a steady state response in ~10 min which was due to slow conductivity changes in the phthalocyanine film. The highest sensor response was obtained within a pH range of 5.0–6.5. The measuring range was 0.005 to 0.3 mM with a sensitivity of 0.042 μS/μM of H2O2. The sensor worked continuously for 7 h with more than 30 measurements and had a storage stability of 90 days when stored at 4 °C.
Figure 5. (a) A HRP conductometric sensor for enzymatic detection of H2O2. The sensor was fabricated using HRP as the enzyme and ttb-CUPc as the active/sensitive film to detect the released iodine molecule. Reprinted from [38] with permission from Elsevier. (b) A bi-enzyme sensor (Alcohol oxidase and catalase) to detect alcohol using conductometric transducer. Reprinted from [39] with permission from Elsevier. (c) (i) Schematics of a conductometric transducer working as an immunosensor (ii) A nanogold particle coated with anti-HBs and protein A (iii) Schematics of the sandwich immunoassay with the reactions involved to measure Hepatitis-B surface antigens. Reprinted from [42] with permission from Elsevier. (d) Schematics of a GOD immobilized cellulose-SnO2 composite electrode. Reprinted from [46] with permission from Elsevier.
Since then, multiple conductometric sensors have been reported in the literature. These sensors can be classified based on electrode material used for the sensors such as metal [43][54][55][56], metal nanoparticles [21][22][23][57], and others [58]. The sensors are compared based on crucial parameters like sensitivity, measuring range, LOD, potential bias, and response time. A summary of conductometric sensors is given in Table 2.
Table 2. Summary of H2O2 conductometric sensors with the crucial sensor properties including LOD, measuring range, voltage bias, response time, buffer and working pH. Where AOX: Alcohol oxidase, PVA: Polyvinyl alcohol, g-C3N4: graphitic carbon nitride, Au: Gold, AuNPs: Gold nanoparticles, AFP: alpha-fetoprotein, LOD: Lactate oxidase.
Substrate |
Target Analyte |
Ligand/ Enzyme |
LOD (µM) |
Measuring Range (mM) |
Voltage Bias (mV) (Frequency) |
Response Time (Minutes) |
Buffer/Working pH |
Comments |
Interference Tested |
Ref |
Metal interdigitated electrodes |
||||||||||
Ceramic-Au |
H2O2 |
Pthalocyanine |
NR |
0.005–0.3 |
60 |
10 |
Working pH 6.0 Buffer: Phosphate (20 mM) |
Storage stability for 90 days at 4 °C |
No |
[43] |
Silicon-Au |
H2O2/ Cyanide |
PVA-Catalase |
6 |
0–100 |
10 (100 kHz) |
5 |
Working pH 7.2 Buffer: Phosphate (5 mM) |
Inhibitory assay for cyanide detection |
No |
[55] |
Au |
Methanol |
AOX-Catalase |
0.5 |
<0.075 |
10 (100 kHz) |
<10 |
Working pH 7.2 Buffer: Phosphate (5 mM) |
Alcoholic beverages |
Yes |
[54] |
Ethanol |
1 |
<0.070 |
||||||||
Propanol |
3 |
<0.065 |
||||||||
Ceramic-Au |
Lactate |
LOD-HRP |
0.05 |
0–0.21 |
10 (100 kHz) |
~20 (est.) |
Working pH 6 Buffer: Phosphate (5 mM) |
Diluted yogurt samples Storage stability for 40 days at 4 °C |
Yes |
[56] |
Metal nanoparticles |
||||||||||
AuNPs |
Hepatitis B (HB) |
HRP/Anti-HBs |
0.01 ng/mL |
0.1–600 ng/mL |
10 (100 kHz) |
>30 |
Working pH 7.0 Buffer: Phosphate (10 mM) |
Tested with serum samples Assay stable for 16 days when stored at 4 °C |
Yes |
[57] |
Ceramic-Au & magnetic NPs |
Glucose |
GOD |
3 |
0.04–3 |
10 (100 kHz) |
<10 |
Working pH 7.3 Buffer: Phosphate (5 mM) |
Stable for 12 days when stored at 4 °C |
No |
[23] |
g-C3N4 |
AFP/H2O2 |
Pt NPs |
0.01 ng/mL |
0.01–100 ng/mL |
10 (100 kHz) |
5–6 |
Working pH 6.5 Buffer: PBS (10 mM) |
Tested with human serum Inhibitory Immunoassay |
Yes |
[21] |
Others |
||||||||||
Cellulose- SnO2 |
H2O2/ Glucose |
GOD |
500 |
0.5–12 |
0–3 V (dc) |
NR |
Working pH 7.2 Buffer: Phosphate |
Storage stability > 10 days Flexible one-time use sensor |
No |
[58] |
H2O2 is not an ideal molecule to detect using electrochemical reactions due to its slow reaction kinetics. Amperometric sensors overcome this by applying a suitable potential to accelerate the reaction kinetics. However, this overpotential limits the detection at low concentrations because of a high signal to noise ratio. Similar detection limit issues exist for potentiometric sensors. In FET sensors, the phenomenon of surface charging at the gate and the effect of it on the inversion layer created and the conductivity of the channel is non-linear and hence the amplification can result. FET sensors can be classified in to four major groups based on the active material used for the detection: silicon nitride [40][41][42][45][46][59][60][61], conducting polymers [62][63][64][65][66], metal oxide [44][67][68][69][70][71][72][73], and carbon nanomaterials [24][25][74][75][76][77][78]. The sensors are evaluated based on crucial parameters like measuring range, sensitivity, detection limit, and response time. FET sensors are summarized in Table 3.
Table 3. Summary of various FETs used for H2O2 measurement where OC: operating conditions; PANI: Polyaniline; PAA: poly acrylic acid; pDAB: poly(1,2 diaminobenzene); HRP: Horseradish peroxidase; PVP: poly(4-vinylpyridine-co-styrene); MOSC: Metal oxide semiconductor capacitor; 3-aminopropyltriethoxysilane (APTES), PPyNT: Polypyrrole nanotube, Pt: Platinum, Is: source current, Vds: Drain potential with respect to source, Vg: is gate voltage, Ids: current between drain and source, Vd: Drain potential, Vbias: potential bias applied between the Pt electrode and reference electrode, I bias: current bias, pDAB: poly(1,2-diaminoben-zene), ITO: Indium titanium oxide, GluD: Glutamate dehydrogenase, UA: Uric acid, AA: Ascorbic acid and MoS2: Molybdenum disulphide.
Substrate |
Target Analyte |
Ligand/ Enzyme |
LOD (µM) |
Sensitivity |
Measuring Range (mM) |
OC |
Response Time (s) |
Working pH & Buffer |
Comments |
Ref |
Silicon nitride FET |
||||||||||
Si3N4-FET |
H2O2 |
HRP |
5 |
~15 mV/mM (est.) |
<2 |
Is: 300 μA Vds: 2 V |
30–90 |
Buffer: Phosphate (10 mM) Working pH: 6 |
<10% reduction in enzyme activity after 1000 measurements |
[41] |
Si3N4-FET/Pt electrode |
Glucose |
GOD |
NR |
~40 mV/mM |
<5 |
Vbias: 0.64 V |
~480 |
Buffer: Phosphate (5–20 mM) Working pH: 7.4 |
Baseline established by removing the potential bias |
[40] |
Si3N4-FET/Pt electrode |
Glucose & sucrose |
GOD & Invertase-mutarotase-GOD |
~50 (est) |
NR |
1.67–16.67 |
Vbias: 0.7 V |
180–300 |
Buffer: Phosphate (10 mM) Working pH: 7.4 |
Greater Pt area, increases sensitivity |
[42] |
Si3N4-FET/Pt electrode |
Glucose |
GOD |
1000 (est.) |
~11 mV/mM (est.) |
1–10 |
Vbias: 0.7 V |
~60 |
Buffer: Phosphate (10 mM) Working pH: 7.4 |
Ladder shape Pt electrode was used for potential bias |
[45] |
Si3N4-FET |
H2O2 |
Pt |
10000 |
5 mV/mM |
10–100 |
Ids: 0.1 mA Vds: IV |
300 |
Buffer: Phosphate (100 mM) Working pH: 7.2 |
Used for glucose and lactate |
[46] |
Conducting polymers |
||||||||||
Carbon |
H2O2 |
PANI-pDAB-HRP |
100 |
NR |
<0.5 |
Vg: 200 mV Vd: 20 mV |
~100 s |
Buffer: citrate-phosphate-Na2SO4 Working pH: 5 |
HRP inhibition at H2O2 concentration > 0.5 mM. |
[74] |
Kapton-Carbon |
H2O2 |
PSPANI-HRP |
25 |
0.126 μA/s |
0.025–1 |
Vg: 0 V Vd: 20 mV |
100–300 s |
Buffer: HEPES-KNO3 (100 mM) Working pH: 7 |
Sultonation improves the PANI conductivity at pH 7 |
[75] |
Si3N4-FET |
Glucose |
PANI-PAA-GOD |
NR |
1 nA/mM |
0–9 |
Vg: 20 mV Vds: 10 mV |
<1 s |
Buffer: McIlvaine Working pH: 5 |
PANI-PAA film was deposition by electropolymerization |
[60] |
PEDOT-TFT |
H2O2 & Glucose |
GOD |
100 |
NR |
0.1–1 |
Vd: 0.2 V Vg: 0–0.6 V |
~60 s |
Buffer: PBS Working pH: 7.14 |
pH independent response from pH 5 to 9 |
[79] |
PEDOT-TFT |
Glucose |
GOD |
1 |
0.1 V/decade |
<1 |
Vds: −0.2 V |
NR |
Buffer: PBS (15 mM) Working pH: 6.8 |
Sensitivity can be improved by increasing Vg |
[62] |
PEDOT-TFT |
Glucose |
GOD |
<1000 |
1.65 μA/mM |
1.1 to 16.5 |
Vds: −1.5 V Vg: 0.0 V |
10–20 s |
NR |
Sensor was encapsulated in cellulose acetate membrane |
[63] |
Liquid gate-FET |
H2O2 & Glucose |
PPyNT-GOD |
500 (est.) |
3.75%/mM (est.) |
2–20 |
Vds: −0.01 V Vg: 0.01 V |
5–10 s |
Buffer: PBS (10 mM) Working pH: 7.0 |
High enzyme loading was achieved |
[64] |
TFT |
H2O2 & Glucose |
PEDOT-GOD |
1 |
0.79–3 μA/mM |
0.001–5 |
Vds: −0.4 V Vg: 0.4 V |
<20 s |
Buffer: PBS Working pH: 7.4 |
Used as both optical and electrochemical |
[80] |
TFT |
Glucose |
PEDOT-GOD |
10 |
NR |
0.01–100 |
Vds: −0.7 V Vg: 0.7 V |
~360 s |
Buffer: PBS (120 mM) |
Stable for 100 days with covalently immobilized GOD |
[81] |
TFT |
H2O2 & Glucose |
PEDOT-TiO2-GOD |
1 |
0.126%/decade |
0.001–5 |
Vds: −0.1 V Vg: 0.4 V |
~1000 s (est.) |
Buffer: PBS (10 mM) Working pH: 7.0 |
Stable for 10 days with intermittent testing |
[82] |
Liquid gate FET |
H2O2 |
rGO-PPy NTs |
0.1 nM |
2%/decade |
0.1–100 nM |
Vg: 0.1 V Vds: −0.01 V |
<1 s |
Buffer: PBS Working pH: 7.4 |
Stable up to 1 month, when stored in air |
|
Metal oxides |
||||||||||
Glass-ITO-SnO2 |
Glucose |
GOD-MnO2 |
2700 |
2.35 mV/mM |
<20 |
No bias |
720 |
Buffer: Phosphate-KOH (5 mM) Working pH: 8.1 |
Dynamic range strongly depends on pH value |
[67] |
Si3N4-FET |
Glucose |
GOD-MnO2 NPs |
20 |
NR |
0.025–1.9 |
No bias |
~140 s |
Buffer: Tris (10 mM) Working pH: 7.4 |
Repeatability: 1.9% (RSD) for 7 measurements |
[61] |
FET |
H2O2 |
Iridium oxide |
100 |
400 mV/dec |
0.1–10 |
Ibias: 25 nA |
NR |
Working pH: 3.5–9 |
- |
[68] |
Prussian blue |
10 |
290 mV/dec |
0.01–1 |
Ibias: 50 nA |
Working pH: 4.5–6 |
|||||
Os-PVP-HRP |
0.1 |
700 mV/dec |
10−7–10−5 M |
Ibias: 25 nA |
Working pH: 4.5–6 |
|||||
Ta2O5-FET-Pt |
H2O2 |
Perovskite oxide |
4 |
35 mV/dec |
0.005–0.2 |
Ibias: 25 nA |
1800 |
Buffer: Phosphate Working pH: 7 |
Change in stoichiometry of oxide can result in lower detection limit |
[44] |
FET |
H2O2 |
TiO2 |
NR |
4.5 mV/μM (DMEM media) |
NR |
Ids: 0.1 mA Vds: 1 V |
300 (est.) |
Buffer: Phosphate |
DMEM media |
[70] |
FET |
Glucose |
ZnO-NiO quantum dots |
26 |
13.14 μA mM−1(0.001–10 mM) |
0.001–50 |
Vg: 1.2–2 V Vds: 0.0 V |
NR |
Buffer: PBS (10 mM) Working pH: 7.4 |
Tested in whole blood and serum |
[83] |
Liquid gate FET |
Glucose |
ZnO rod-GOD |
0.07 |
32.27 μA mM−1cm−2 |
0.05–70 |
Vg: 0–2 V |
NR |
Buffer: PBS (50 mM) Working pH: 7.4 |
Mice blood, serum |
[73] |
Cholesterol |
ZnO rod-COD |
0.04 |
17.1 μA mM−1cm−2 |
0.01–45 |
Vg: 2–3 V |
NR |
||||
Carbon nanomaterials |
||||||||||
Graphene-FET |
Glucose |
GOD |
100 |
~1 μA/mM (est.) |
<10 |
Vds: 0.1 V Vg: 0 V |
<200 s (est.) |
Buffer: PBS (10 mM) Working pH: 7.2 |
Glutamate was also detected using the sensor with GluD |
[76] |
OTFT |
Glucose |
Graphene-Chitosan-GOD |
0.01 |
370 mV/dec |
0.01–1μM |
Vg: 0.4 V Vds: 0.05 V |
~500 s |
Buffer: PBS Working pH: 7.4 |
Investigated the effect of interference of UA and AA |
[84] |
Graphene-FET |
Glucose |
Silk fibroin-GOD |
100 |
2.5 μA/mM |
0.1–10 |
Vg: 0 V Vds:0.1 V |
~100 s |
Buffer: PBS (10 mM) Working pH: 7.4 |
Stable for 10 months at room temperature |
[78] |
FET |
H2O2 & Glucose |
Graphene-Chitosan-PtNPs-GOD |
0.03 |
91.7 mV/dec |
30 nM–1 mM |
Vg: 0.7 V Vds: 0.05 V |
~100 s (est.) |
Buffer: PBS Working pH: 7.2 |
No interference was observed from AA and UA |
[85] |
rGO-FET |
H2O2 |
MoS2 |
1 pM |
0.46%/dec |
1 pM–100 nM |
Vg: 0.1 V Vds: 0.01 V |
~1 s |
Buffer: PBS Working pH: 7.4
|
HeLa Cells |
[25] |
FET |
H2O2 |
Graphene-Cyt-c |
0.1 pM |
14%/dec |
0.1–100 pM |
Vg: 1.75 V Vds: 0.001 V |
<1 s |
Buffer: PBS Working pH: 7.4 |
No interference from UA, AA, dopamine, and glutamate |
[24] |
Others |
||||||||||
SiO2-MOSC |
Glucose |
HRP-GOD |
5000 |
1.76 nA/cm2M |
<2 M |
Vg: 5 V |
1200 |
Dry sensor so no need for a buffer solution |
- |
[69] |
Polysilicon wire-ISFET |
H2O2 & Glucose |
APTES-SiNPs-UV treatment |
32 pM |
12 AmM−1cm−2 |
10−10–10−3 M |
Vds: 5 V |
NR |
Tested solution volume: 0.03 pL (Dry sensor) |
Serum |
[72] |