Evolution of Fetal Cardiac Imaging: History
Please note this is an old version of this entry, which may differ significantly from the current revision.
The evaluation of a variety of congenital heart diseases (CHDs) using three-dimensional (3D) ultrasound with different software, such as Cristal Vue, Realistic Vue, LumiFlow, and Spatiotemporal Image Correlation (STIC), with HDlive and HDlive Flow Silhouette modes. These technologies provide realistic images of the fetal heart and cardiac vessels using a fixed virtual light source that allows the operator to freely select a better light source position to enhance the cardiovascular anatomical details. In addition, Fetal Intelligent Navigation Echocardiography (FINE) technology, also known as “5D Heart” or “5D”, is a technology that enables the automatic reconstruction of the nine standard fetal echocardiographic views and can alert non-specialists to suspected CHD. Through the use of artificial intelligence, an ultrasound machine is able to perform automatic anatomical and functional measurements. In addition, these technologies enable the reconstruction of fetal cardiac structures in realistic images, improving the depth perception and resolution of anatomic cardiac details and blood vessels compared to those of standard two-dimensional (2D) ultrasound.
  • fetal heart
  • spatiotemporal image correlation
  • HDlive
  • fetal intelligent navigation echocardiography
  • artificial intelligence

1. Introduction

Congenital heart disease (CHD) is the most common birth defect and a leading cause of perinatal mortality related to birth defects [1]. The prenatal diagnosis of cardiac anomalies is a key element in improving the prognosis of structural CHD and optimizing the prevention of cardiovascular disease in hearts that have suffered a prenatal insult, such as fetuses with growth restriction. In this scenario, the development of prenatal cardiac ultrasound screening programs has improved the early detection of CHD. The detection rates have been improved by adding ventricular outflow tracts and superior mediastinal views to the classic standard four-chamber (4C) view during cardiac ultrasound [2].
The advent of advanced imaging technologies, such as three-dimensional (3D) ultrasound in the early 1990s and, in particular, Spatiotemporal Image Correlation (STIC) software in the early 2000s, enabled major technological advances in the 3D assessment of the fetal heart. STIC technology, as first described by De Vore et al. [3], involves the acquisition of approximately 150 two-dimensional images per second using a volumetric transducer during a single 7.5 to 15 s scan. The acquired 3D volumes contain a “block” of cardiac ultrasound images that allow the detailed analysis of cardiac anatomy [4]. This technology also provides measurements of the areas of the fetal heart valves and the volumes of the chambers and myocardial walls. Some of the advantages of STIC in the assessment of the fetal heart include less operator dependence in obtaining ultrasound views of the heart, a shorter scan examination time, as the volume analysis can be performed in the absence of the pregnant woman (offline), and the ability to send the heart volumes via an Internet link for review by experts in tertiary centers (tele-STIC) [5][6].
In this scenario, new software called HDlive (General Electric Healthcare, Zipf, Austria) can be added to STIC to improve the quality of the ultrasound images. This technology differs from conventional 3D rendering methods in that the operator can freely select the light source at any angle, providing realistic sensations of the cardiac structures and their vessels, increasing the accuracy of cardiac anatomy detailing [7][8][9][10].
Fetal Intelligent Navigation Echocardiography (FINE), also known as “5D Heart” or “5D”, presents nine standard fetal echocardiographic views by performing cardiac volume acquisition from a 4C chamber view of the fetal heart [11]. Currently, with the advent of “artificial intelligence”, the addition of this technology to 5D Heart can provide alerts for possible CHD in the images obtained via automatic reconstructions of the cardiac ultrasound views. Artificial intelligence is a promising technological innovation that can alert non-specialists to the suspicion of CHD, provide automatic measurements of image acquisition, and reduce the differences between inter-observer measurements.

2. Three-Dimensional Ultrasound with Spatiotemporal Image Correlation

Three-dimensional (3D) ultrasound with Spatiotemporal Image Correlation (STIC) was introduced in the early 2000s. This technology allows the volumetric acquisition of the fetal heart and its vessels, offering the ability to evaluate them in multiplanar and rendering modes, either static or in motion (four-dimensional—4D), reproducing a complete cardiac cycle [12][13].
The advantages of STIC are its higher sensitivity in the diagnosis of CHD, especially in conotruncal anomalies, and the shorter time of in-person ultrasound examination with the possible offline analysis of the fetal heart (after the pregnant woman has left the examination room). The details of morphology and function can be assessed using the rendering mode and, in complex cases, cardiac volumes can be sent via the Internet to reference centers specializing in fetal echocardiography. This technique is therefore suitable for ultrasound cardiac screening programs [14][15][16].
To obtain a cardiac volume using the STIC method, the examiner must obtain a 4C view of the fetal heart, whenever possible, with the fetal spine positioned as close to 6 o’clock as possible. Ideally, this image acquisition should be performed with the pregnant woman in apnea for a few seconds and the fetus at rest. The settings for this acquisition are: an opening angle between 20° and 40° and an acquisition time between 7.5 and 15 s according the gestational age [17]. By adjusting the brightness and color, STIC in the rendering mode allows the detailed evaluation of cardiac anatomy and measurement (area and volume) of a variety of structures of the fetal heart, such as the interventricular septum, valves, and papillary muscles [18][19][20][21][22]. STIC technology can reconstruct moving images and display them in grayscale, Doppler mode (inversion mode/color Doppler), B-flow imaging, and Tomographic Ultrasound Imaging (TUI) [23][24][25][26].
Grayscale STIC allows the assessment of the ventricular outflow tracts as well as the aortic and ductal arches. In this scenario, the B-flow imaging technique in STIC enhances the weak signals from blood and eliminates strong signals from nearby structures and is potentially superior to color Doppler in the assessment of large vessels and venous connections. B-flow imaging produces images similar to angiographic images. The post-processing inversion mode (inverted flow) is another technique that analyzes fluid structures and inverts the grayscale voxels. The inversion mode allows the reconstruction of ventricles, aortic, and ductal arches, as well as abnormal venous connections and septal defects.
Fetal cardiac volumes and functional parameters (stroke volume, cardiac output, and ejection fraction) can be assessed using STIC with the Virtual Organ Computer-Assisted Analysis (VOCAL) method. Each ventricular volume is obtained at the end of the diastole (maximum volume) and at the end of the systole (minimum volume), with the caliper positioned on the inner side of the basal and apical ventricular walls. After six sequential views, which are manually delineated for each ventricle, VOCAL provides the chamber volume and reconstructs its 3D image. Finally, the stroke volume, cardiac output, and ejection fraction of each ventricle are calculated from these measurements [27][28][29][30].

3. HDlive

The new applications, HDlive, HDlive Flow, and HDlive Silhouette mode (General Electric Healthcare, Zipf, Austria), with STIC have the potential to provide a more realistic appearance of the cardiac structures and demonstrate blood flow with vitreous-like clarity. The HDlive Flow Silhouette mode has the ability to outline the walls of blood vessels, while showing the vessel lumen as semi-transparent. This technology demonstrates blood flow with a clarity similar to that of the vitreous. The accuracy of this technique can assist the team of professionals involved in the ultrasound assessment of the fetal heart by providing a better delineation of the blood vessels with a positive impact on the prenatal diagnosis of CHD [7][8][9][10][31][32]. HDlive Flow uses an examiner-adjustable light source to create light and shadow effects, increasing the depth perception of the 3D/4D color/power Doppler ultrasound. Hata et al. [33] demonstrated that the resolution and image quality of HDlive Flow are significantly higher than those obtained from conventional 3D/4D power Doppler, which may provide a better understanding of fetal cardiac anatomy.
Recently, Li et al. [34] demonstrated, in a retrospective study of 174 fetuses with conotruncal anomalies, that the accuracy of STIC with HDlive increases the detection of these cardiac anomalies, especially truncus arteriosus. The conotruncal anomalies detected in this study were described as follows: tetralogy of Fallot (58 cases), the transposition of the great arteries (30 cases), the double outlet of the right ventricle (26 cases), and 32 cases of truncus arteriosus. Complex CHD technology provides high-quality images that detail the anatomy, increasing the diagnostic accuracy [35].

4. Fetal Intelligent Navigation Echocardiography “5D Heart”

The Fetal Intelligent Navigation Echocardiography (FINE) technique, known as “5D Heart” (Samsung Healthcare, Gangwon, South Korea), consists of the automatic reconstruction of the nine standardized fetal echocardiographic landmarks after marking strategic cardiac anatomical points from the four-chamber view of the fetal heart. Using intelligent navigation, this software guides seven points to be marked by the examiner (the descending aorta, crux cordis, pulmonary valve, superior vena cava, and transverse aorta). The nine views of fetal echocardiography are automatically played in sequence, and the software itself labels the views. It is possible to zoom in on each view separately and to enhance the image separately using the brightness and contrast buttons. By applying FINE technology to STIC volume datasets, the examination of the fetal heart can be simplified with reduced operator dependency [11][36]. By adding color Doppler to FINE, the detection of CHD can be improved by assessing the Doppler flow characteristics of the heart vessels [37]. In a cross-sectional prospective study, Carrilho et al. [36] concluded that the quality of echocardiographic views obtained with the FINE technique was superior to that of the other techniques, such as the simple targeted arterial rendering (STAR) technique and the 4C view swing technique (FAST).

5. Realistic Vue, Crystal Vue, and LumiFlow

Realistic Vue, Crystal Vue, and LumiFlow (Samsung Healthcare, Gangwon, South Korea) provide more detailed and realistic images of the fetus with instant processing, and Crystal Vue improves the visualization of internal and external structures in a single rendered image. LumiFlow is a post-processing and shading technique used to better assess the fetal micro- and macrovasculature of the fetus. This technology can be added to existing Doppler images, such as color/power, to simulate a 3D pace view of the fetal vasculature. With LumiFlow, the center of the vessel is displayed in a slightly lighter color due to its faster velocity, with a darker border around the vessels where blood velocities are slower [38][39].

6. Artificial Intelligence

Congenital heart disease (CHD) can present as a wide variety of cardiac malformations, especially the more complex ones, requiring the team involved in the diagnosis to be well versed in the diversity of cardiac defects. As a result, many CHDs remain undiagnosed during ultrasound. Considering that CHD is a major cause of infant mortality and that prenatal diagnosis is crucial in this regard, the impact of this undiagnosed condition is unfavorable for the survival of these patients. In this scenario, artificial intelligence (AI) can assist non-specialists in diagnosing CHD. Machine learning (ML), which can be defined as the use of computer programs that improve over time, is an integral part of this technology. In addition, Deep Learning (DL) is a specific type of ML that uses neural networks organized into many layers, allowing the interpretation of the data provided, such as image classification in computer vision. As an example, DL is able to automatically detect the prognosis of CHD once the diagnosis is known [40][41].
With ML, the ultrasound machine is able to automatically obtain cardiac measurements based on the identification of anatomical cardiac structures, reducing the scan time and inter-observer measurement variation. In this scenario, not only anatomical, but also functional parameters (such as the myocardial performance index) can be automatically measured. With the advances in AI, this technology has been increasingly added to ultrasound devices, such as obstetric ultrasound, and is applicable to three areas: structure identification, automatic measurements, and the classification of diagnosis. AI software (Heart Assist and MPI+—Samsung Healthcare, Gangwon, South Korea) can identify the fetal heart structures and perform automatic biometric and the right and left myocardial performance index measurements, respectively [42][43][44].

7. Isomerism

The examination of the upper abdomen of the fetus via ultrasound provides a situs assessment. In the transverse abdominal view, if the situs is normal (situs solitus), the aorta is a left-sided vessel and is located posterior to the inferior vena cava. Left atrial isomerism is a condition in which there are two left-sided organs, and right atrial isomerism is the opposite situation (for example, the atria and lungs have the same shape). In left atrial isomerism, the venous vessel is posterior to the aorta due to the absence of the hepatic segment of the inferior vena cava, with drainage via the azygos or hemiazygos veins. The aorta and the azygos or hemiazygos veins are juxtaposed (a “double-vessel” sign) [45][46]. In general, abnormalities are associated with more complex CHDs, such as an unbalanced atrioventricular septal defect [47]. Conversely, in right atrial isomerism, there is a double right-sided atrium. Therefore, in the upper transverse abdominal view, the two vessels (venous and arterial) are located on the right side of the fetal spine. In this scenario, advanced technologies (5D Heart, HDlive, and HDlive Silhouette modes) are useful tools for the detailed anatomical diagnosis of these complex anomalies, as they reduce the ultrasound scan time and provide high-quality images.

8. Atrioventricular Septal Defect

An atrioventricular septal defect (AVSD) is a CHD that results from the failure of the atrioventricular septum to develop, which can lead to abnormalities of the atrioventricular valves, an ostium primum atrial septal defect (ASD), and an inlet ventricular septal defect (VSD). The complete form of an AVSD is characterized by a common atrioventricular valve, a ostium primum-type ASD, and a large inlet ventricular septal defect (VSD). In other forms of AVSDs, the right and left atrioventricular valves are present with a small VSD or intact interventricular septum. The complete form of an AVSD is strongly associated with syndromes such as trisomy 21. Unbalanced complete AVSD (=marked discrepancy between right and left ventricle) is typically associated with heterotaxy syndromes such as left isomerism. The complete form of an AVSD, can be identified on the 4C view of the fetal heart with color Doppler by an “H-shaped” sign due to the absence of the atrioventricular septum. An AVSD may be associated with other cardiac defects such as tetralogy of Fallot, double outlet right ventricle, and anomalies with outflow tract obstruction. Advanced technologies can provide high-quality images to facilitate early detection during the first trimester cardiac screening in cases with increased nuchal translucency and in complex cases, such as it is when associated with other cardiac anomalies [2][3][6][48].

9. Tetralogy of Fallot

Tetralogy of Fallot is the most common conotruncal anomaly and even the most common cyanogenic CHD. Classically, it is characterized by a tetralogy of morphological features, as follows: the obstruction of the right ventricular outflow tract (subpulmonary obstruction), a VSD, the rightward deviation of the aorta (overriding = biventricular connection of the aorta), and right ventricular hypertrophy. In general, right ventricle hypertrophy is not visible in the fetal period, especially with two-dimensional (2D) ultrasound [49]. Right-sided aortic arch (from 13 to 25% of tetralogy of Fallot cases), muscular VSDs, and AVSDs are some of the cardiac anomalies most commonly associated with tetralogy of Fallot. The extra-cardiac defects associated with it are the following: trisomy (13, 18, and 21), CHARGE, VATER, pentalogy of Cantrell, omphalocele, and Di George syndrome (deletion 22q11.2); in the latter one, the evaluation of the thymus is important [50].
For the diagnosis of tetralogy of Fallot, the outflow tract views are critical, and advanced imaging techniques will improve the accuracy and timing of the ultrasound examination, as in classic tetralogy of Fallot, the image of the fetal heart in the 4C view is normal [34][51]. Tetralogy of Fallot with subpulmonary stenosis is the classic form of this CHD, while the extreme form is pulmonary atresia. There is also a rare form of tetralogy of Fallot in which the pulmonary valve is absent [52][53].
The standard 4C view using 2D ultrasound is generally normal in the classic forms of tetralogy of Fallot, where the evaluation of the outflow tract views is critical. Advanced technologies such as 5D Heart can help to reduce the time of the scan examination by allowing the automatic reconstruction of the nine echocardiographic views. Therefore, in the classical forms of tetralogy of Fallot, this diagnosis becomes easier by drawing attention to the small pulmonary artery in the three vessels and trachea view and the overring of the aorta in the left ventricle outflow tract view [51][52]. In extreme cases, the presence of retrograde flow from the ductus arteriosus to the pulmonary artery confirms the atresia, which can be easily detected with advanced technologies. Conversely, in the forms of tetralogy of Fallot in which the pulmonary valve is absent, the right and left pulmonary arteries are enlarged [53].
In fact, the advanced technologies reveal the anatomical details due to high-quality imaging, such as in non-classical forms of tetralogy of Fallot or when it is associated with other heart defects such as an AVSD.

This entry is adapted from the peer-reviewed paper 10.3390/diagnostics13233509

References

  1. Meller, C.H.; Grinenco, S.; Aiello, H.; Córdoba, A.; Sáenz-Tejeira, M.M.; Marantz, P.; Otaño, L. Congenital heart disease, prenatal diagnosis and management. Arch. Argent. Pediatr. 2020, 118, e149–e161.
  2. Carvalho, J.S.; Axt-Fliedner, R.; Chaoui, R.; Copel, J.A.; Cuneo, B.F.; Goff, D.; Kopylov, L.G.; Hecher, K.; Lee, W.; Moon-Grady, A.J.; et al. ISUOG Practice Guidelines (updated): Fetal cardiac screening. Ultrasound Obstet. Gynecol. 2023, 61, 788–803.
  3. DeVore, G.R.; Falkensammer, P.; Sklansky, M.S.; Platt, L.D. Spatio-temporal image correlation (STIC): New technology for evaluation of the fetal heart. Ultrasound Obstet. Gynecol. 2003, 22, 380–387.
  4. Gonçalves, L.F.; Romero, R.; Espinoza, J.; Lee, W.; Treadwell, M.; Chintala, K.; Brandl, H.; Chaiworapongsa, T. Four-dimensional ultrasonography of the fetal heart using color Doppler spatiotemporal image correlation. J. Ultrasound Med. 2004, 23, 473–481.
  5. Viñals, F.; Ascenzo, R.; Naveas, R.; Huggon, I.; Giuliano, A. Fetal echocardiography at 11 + 0 to 13 + 6 weeks using four-dimensional spatiotemporal image correlation telemedicine via an Internet link: A pilot study. Ultrasound Obstet. Gynecol. 2008, 31, 633–638.
  6. Araujo Júnior, E.; Tonni, G.; Bravo-Valenzuela, N.J.; Da Silva Costa, F.; Meagher, S. Assessment of Fetal Congenital Heart Diseases by 4-Dimensional Ultrasound Using Spatiotemporal Image Correlation: Pictorial Review. Ultrasound Q. 2018, 34, 11–17.
  7. Hata, T.; Mashima, M.; Ito, M.; Uketa, E.; Mori, N.; Ishimura, M. Three dimensional HD live rendering images of the fetal heart. Ultrasound Med. Biol. 2013, 39, 1513–1517.
  8. Hata, T.; AboEllail, M.A.M.; Sajapala, S.; Ishimura, M.; Masaoka, H. HDlive silhouette mode with spatiotemporal image correlation for assessment of the fetal heart. J. Ultrasound Med. 2016, 35, 1489–1495.
  9. Hata, T.; Ito, M.; Nitta, E.; Pooh, R.; Sasahara, J.; Inamura, N. HDlive flow silhouette mode for diagnosis of ectopia cordis with a left ventricular diverticulum at 15 weeks’ gestation. J. Ultrasound Med. 2018, 37, 2465–2467.
  10. Hata, T.; Koyanagi, A.; Yamanishi, T.; Bouno, S.; Takayoshi, R.; Miyagi, Y.; Miyake, T. Success rate of five fetal cardiac views using HDlive Flow with spatiotemporal image correlation at 18–21 and 28–31 weeks of gestation. J. Perinat Med 2020, 48, 384–388.
  11. Yeo, L.; Romero, R. Fetal Intelligent Navigation Echocardiography (FINE): A novel method for rapid, simple, and automatic examination of the fetal heart. Ultrasound Obstet. Gynecol. 2013, 42, 268–284.
  12. Yeo, L.; Romero, R.; Jodicke, C.; Kim, S.K.; Gonzalez, J.M.; Oggè, G.; Lee, W.; Kusanovic, J.P.; Vaisbuch, E.; Hassan, S. Simple targeted arterial rendering (STAR) technique: A novel and simple method to visualize the fetal cardiac outflow tracts. Ultrasound Obstet. Gynecol. 2011, 37, 549–556.
  13. Rocha, L.A.; Rolo, L.C.; Barros, F.S.B.; Nardozza, L.M.M.; Moron, A.F.; Júnior, E.A. Assessment of Quality of Fetal Heart Views by 3D/4D Ultrasonography Using Spatio-Temporal Image Correlation in the Second and Third Trimesters of Pregnancy. Echocardiography 2015, 32, 1015–1021.
  14. Espinoza, J.; Lee, W.; Romero, R.; Yeo, L.; Rizzo, G.; Paladini, D.; Viñals, F.; Achiron, R.; Gindes, L.; Abuhamad, A.; et al. Collaborative study on 4-dimensional echocardiography for the diagnosis of fetal heart defects: The COFEHD study. J. Ultrasound Med. 2010, 29, 1573–1580.
  15. Bennasar, M.; Martínez, J.M.; Gómez, O.; Bartrons, J.; Olivella, A.; Puerto, B.; Gratacós, E. Accuracy of four-dimensional spatiotemporal image correlation echocardiography in the prenatal diagnosis of congenital heart defects. Ultrasound Obstet. Gynecol. 2010, 36, 458–464.
  16. Uittenbogaard, L.B.; Haak, M.C.; Spreeuwenberg, M.D.; Van Vugt, J.M.G. A systematic analysis of the feasibility of four-dimensional ultrasound imaging using spatiotemporal image correlation in routine fetal echocardiography. Ultrasound Obstet. Gynecol. 2008, 31, 625–632.
  17. Gonçalves, L.F.; Espinoza, J.; Lee, W.; Nien, J.K.; Hong, J.S.; Santolaya-Forgas, J.; Mazor, M.; Romero, R. A new approach to fetal echocardiography: Digital casts of the fetal cardiac chambers and great vessels for detection of congenital heart disease. J. Ultrasound Med. 2005, 24, 415–424.
  18. Rolo, L.C.; Santana, E.F.M.; da Silva, P.H.; Costa, F.D.S.; Nardozza, L.M.M.; Tonni, G.; Moron, A.F.; Júnior, E.A. Fetal cardiac interventricular septum: Volume assessment by 3D/4D ultrasound using spatio-temporal image correlation (STIC) and virtual organ computer-aided analysis (VOCAL). J. Matern. Fetal Neonatal Med. 2015, 28, 1388–1393.
  19. Xiong, Y.; Liu, T.; Wu, Y.; Xu, J.F.; Ting, Y.H.; Yeung Leung, T.; Lau, T.K. Comparison of real-time three-dimensional echocardiography and spatiotemporal image correlation in assessment of fetal interventricular septum. J. Matern. Fetal Neonatal Med. 2012, 25, 2333–2338.
  20. Adriaanse, B.M.; Uittenbogaard, L.B.; Tromp, C.H.; Heymans, M.W.; van Vugt, J.M.; Haak, M.C. Prenatal Visualization of the Pulmonary and Aortic Valves and Leaflets Is Feasible Using 4-Dimensional Sonography. J. Ultrasound Med. 2016, 35, 497–504.
  21. Rolo, L.C.; Rizzo, G.; Pietrolucci, M.E.; Barros, F.S.B.; Nardozza, L.M.M.; Martins, W.P.; Arduini, D.; Moron, A.F.; Júnior, E.A. Viewing rate and reference ranges for papillary muscle areas of the fetal heart using four-dimensional ultrasound in the rendering mode. Prenat. Diagn. 2014, 34, 1153–1160.
  22. Rolo, L.C.; Pietrolucci, M.E.; Araujo Júnior, E.; Barros, F.S.B.; Nardozza, L.M.M.; Martins, W.P.; Moron, A.F.; Arduin, D.; Rizzo, G. Viewing rate and reproducibility of papillary muscle areas in foetal atrioventricular valves using spatio-temporal image correlation in the rendering mode in congenital heart disease. J. Matern. Neonatal Med. 2015, 28, 1375–1380.
  23. Lima, A.I.F.; Júnior, E.A.; Martins, W.P.; Nardozza, L.M.M.; Moron, A.F.; Pares, D.B.S. Assessment of the fetal heart at 12–14 weeks of pregnancy using B-mode, color Doppler, and spatiotemporal image correlation via abdominal and vaginal ultrasonography. Pediatr. Cardiol. 2013, 34, 1577–1582.
  24. Lei, W.; Ying, Z.; Ailu, C.; Xiaoguang, W. Evaluation of normal fetal ductus venosus using B-flow imaging with spatiotemporal image correlation and traditional color Doppler echocardiography. Echocardiography 2015, 32, 325–331.
  25. Qin, Y.; Zhang, Y.; Zhou, X.; Wang, Y.; Sun, W.; Chen, L.; Zhao, D.; Cai, A. Four-dimensional echocardiography with spatiotemporal image correlation and inversion mode for detection of congenital heart disease. Ultrasound Med. Biol. 2014, 40, 1434–1441.
  26. Rizzo, G.; Capponi, A.; Vendola, M.; Pietrolucci, M.E.; Arduini, D. Role of tomographic ultrasound imaging with spatiotemporal image correlation for identifying fetal ventricular septal defects. J. Ultrasound Med. 2008, 27, 1071–1075.
  27. Messing, B.; Cohen, S.M.; Valsky, D.V.; Rosenak, D.; Hochner-Celnikier, D.; Savchev, S.; Yagel, S. Fetal cardiac ventricle volumetry in the second half of gestation assessed by 4D ultrasound using STIC combined with inversion mode. Ultrasound Obstet. Gynecol. 2007, 30, 142–151.
  28. DeKoninck, P.; Steenhaut, P.; Van Mieghem, T.; Mhallem, M.; Richter, J.; Bernard, P.; De Catte, L.; Deprest, J. Comparison of Doppler-based and three-dimensional methods for fetal cardiac output measurement. Fetal Diagn Ther. 2012, 32, 72–78.
  29. Bravo-Valenzuela, N.J.; Peixoto, A.B.; Mattar, R.; Júnior, J.F.M.; Pares, D.B.d.S.; Júnior, E.A. Fetal Cardiac Function and Ventricular Volumes Determined by Three-Dimensional Ultrasound Using STIC and VOCAL Methods in Fetuses from Pre-gestational Diabetic Women. Pediatr. Cardiol. 2020, 41, 1125–1134.
  30. Júnior, J.F.M.; Bravo-Valenzuela, N.J.; Nardozza, L.M.M.; Peixoto, A.B.; Mattar, R.; Martins, W.P.; Pares, D.B.d.S.; Júnior, E.A. References Values of Fetal Heart Myocardial Volume by Three-Dimensional Ultrasound using Spatiotemporal Image Correlation and Virtual Organ Computer-Aided Analysis Methods and Their Applicability in Pregestational Diabetic Women. Am. J. Perinatol. 2021, 38, 721–727.
  31. Ito, M.; AboEllail, M.A.M.; Yamamoto, K.; Kanenishi, K.; Tanaka, H.; Masaoka, H.; Hata, T. HDlive Flow silhouette mode and spatiotemporal image correlation for diagnosing congenital heart disease. Ultrasound Obstet. Gynecol. 2017, 50, 411–415.
  32. Malho, A.S.; Bravo-Valenzuela, N.J.; Ximenes, R.; Peixoto, A.B.; Araujo Júnior, E. Antenatal diagnosis of congenital heart disease by 3D ultrasonography using spatiotemporal image correlation with HDlive Flow and HDlive Flow silhouette rendering modes. Ultrasonography 2022, 41, 578–596.
  33. Hata, T.; AboEllail, M.A.M.; Sajapala, S.; Ito, M. HDliveFlow in the assessment of fetal circulation. Donald School J. Ultrasound Obstet. Gynecol. 2015, 9, 462–470.
  34. Li, T.G.; Su, X.R.; Wu, W.R.; Zhang, W.D.; Ma, B. Prenatal diagnosis of fetal conotruncal defects by using 2D ultrasound and HD live flow combined with spatiotemporal image correlation. J. Clin. Ultrasound 2023, 51, 1166–1171.
  35. Zhang, Y.; Shao, Q.; Sun, L.; Ge, S.; He, Y. Prenatal diagnosis of total anomalous pulmonary venous connection by 2D and 3D fetal echocardiography. Echocardiography 2017, 34, 1852–1857.
  36. Carrilho, M.C.; Rolo, L.C.; Tonni, G.; Araujo Júnior, E. Assessment of the quality of fetal heart standard views using the FAST, STAR, and FINE four-dimensional ultrasound techniques in the screening of congenital heart diseases. Echocardiography 2020, 37, 114–123.
  37. Yeo, L.; Romero, R. Color and power Doppler combined with Fetal Intelligent Navigation Echocardiography (FINE) to evaluate the fetal heart. Ultrasound Obstet. Gynecol. 2017, 50, 476–491.
  38. Pappalardo, E.; Gulino, F.A.; Ettore, C.; Cannone, F.; Ettore, G. Body Stalk Anomaly Complicated by Ectopia Cordis: First-Trimester Diagnosis of Two Cases Using 2- and 3-Dimensional Sonography. J. Clin. Med. 2023, 12, 1896.
  39. Dall’Asta, A.; Grisolia, G.; Volpe, N.; Schera, G.; Sorrentino, F.; Frusca, T.; Ghi, T. Prenatal visualization of the torcular herophili by means of a Doppler technology highly sensitive for low-velocity flow in the expert assessment of the posterior fossa: A prospective study. BJOG 2021, 128, 347–352.
  40. Day, T.G.; Kainz, B.; Hajnal, J.; Razavi, R.; Simpson, J.M. Artificial intelligence, fetal echocardiography, and congenital heart disease. Prenat. Diagn. 2021, 41, 733–742.
  41. Rizzo, G.; Pietrolucci, M.E.; Capponi, A.; Mappa, I. Exploring the role of artificial intelligence in the study of fetal heart. Int. J. Cardiovasc. Imaging 2022, 38, 1017–1019.
  42. Wu, H.; Wu, B.; Lai, F.; Liu, P.; Lyu, G.; He, S.; Dai, J. Application of Artificial Intelligence in Anatomical Structure Recognition of Standard Section of Fetal Heart. Comput. Math. Methods Med. 2023, 2023, 5650378.
  43. Truong, V.T.; Nguyen, B.P.; Nguyen-Vo, T.-H.; Mazur, W.; Chung, E.S.; Palmer, C.; Tretter, J.T.; Alsaied, T.; Pham, V.T.; Do, H.Q.; et al. Application of machine learning in screening for congenital heart diseases using fetal echocardiography. Int. J. Cardiovasc. Imaging 2022, 38, 1007–1015.
  44. Scharf, J.L.; Dracopoulos, C.; Gembicki, M.; Welp, A.; Weichert, J. How Automated Techniques Ease Functional Assessment of the Fetal Heart: Applicability of MPI+™ for Direct Quantification of the Modified Myocardial Performance Index. Diagnostics 2023, 13, 1705.
  45. Sheley, R.C.; A Nyberg, D.; Kapur, R. Azygous continuation of the interrupted inferior vena cava: A clue to prenatal diagnosis of the cardiosplenic syndromes. J. Ultrasound Med. 1995, 14, 381–387.
  46. Bronshtein, M.; Gover, A.; Zimmer, E.Z. Sonographic definition of the fetal situs. Obstet. Gynecol. 2002, 99, 1129–1130.
  47. Anderson, R.H.; Shirali, G. Sequential segmental analysis. Ann. Pediatr. Cardiol. 2009, 2, 24–35.
  48. Veronese, P.; Guariento, A.; Cattapan, C.; Fedrigo, M.; Gervasi, M.T.; Angelini, A.; Riva, A.; Vida, V. Prenatal Diagnosis and Fetopsy Validation of Complete Atrioventricular Septal Defects Using the Fetal Intelligent Navigation Echocardiography Method. Diagnostics 2023, 13, 456.
  49. Anderson, R.H.; Spicer, D.E.; Giroud, J.M.; Mohun, T.J. Tetralogy of Fallot: Nosological, morphological, and morphogenetic considerations. Cardiol. Young 2013, 23, 858–866.
  50. Sharma, R.; Niederhoffer, K.Y.; Caluseriu, O.; Cooke, C.; Hornberger, L.K.; He, R.; Eckersley, L.; Lin, L.; Rushfeldt, M.; McBrien, A. Extra-cardiac diagnoses and postnatal outcomes of fetal tetralogy of fallot. Prenat. Diagn. 2022, 42, 260–266.
  51. Yoo, S.J.; Lee, Y.-H.; Kim, E.S.; Ryu, H.M.; Kim, M.Y.; Choi, H.-K.; Cho, K.S.; Kim, A. Three-vessel view of the fetal upper mediastinum: An easy means of detecting abnormalities of the ventricular outflow tracts and great arteries during obstetric screening. Ultrasound Obstet. Gynecol. 1997, 9, 173–182.
  52. Martínez, J.M.; Gómez, O.; Bennasar, M.; Olivella, A.; Crispi, F.; Puerto, B.; Gratacós, E. The ‘question mark’ sign as a new ultrasound marker of tetralogy of Fallot in the fetus. Ultrasound Obstet. Gynecol. 2010, 36, 556–560.
  53. Gottschalk, I.; Jehle, C.; Herberg, U.; Breuer, J.; Brockmeier, K.; Bennink, G.; Hellmund, A.; Strizek, B.; Gembruch, U.; Geipel, A.; et al. Prenatal diagnosis of absent pulmonary valve syndrome from first trimester onwards: Novel insights into pathophysiology, associated conditions and outcome. Ultrasound Obstet. Gynecol. 2017, 49, 637–642.