After sample collection, extraction techniques are employed to isolate the pesticide compounds from the collected air samples. Common extraction methods include solid-phase extraction (SPE), liquid-liquid extraction (LLE), and solid-phase microextraction (SPME). With the increasing focus on miniaturization and the adoption of green chemistry principles, environmentally friendly sample preparation methods have gained popularity for pesticide extraction. One such method is pressurized liquid extraction (PLE), which offers high efficiency and low consumption of organic solvents for extracting pesticides from particulate matter (PM)
[9]. Microwave-assisted extraction (MAE) is another technique that has been utilized for pesticide extraction, offering similar advantages
[10]. In contrast, miniaturized systems based on ultrasonic-assisted extraction (UAE) have emerged as a simple and cost-effective alternative for extracting pollutants from PM
[11]. Moreover, different solvents and sorbents can be utilized depending on the properties of the target pesticides. Analytical methods such as gas chromatography (GC), liquid chromatography (LC), or their combination with mass spectrometry (MS) are commonly employed for the analysis and quantification of airborne pesticides
[12]. In addition to the traditional methods, alternative techniques like ion mobility spectrometry (IMS) and attenuated total reflectance–Fourier-transform infrared spectroscopy (ATR-FTIR) have been applied to measure airborne pesticides
[13][14][13,14]. These methods allow for the identification and quantification of specific pesticide compounds present in the samples. Additionally, high-resolution mass spectrometry (HRMS) techniques offer improved sensitivity and selectivity in pesticide analysis
[15].
2. Pesticide Concentrations in Air
2.1. Indoor Air
The potential adverse health effects stemming from pesticide exposure have garnered substantial public attention. While numerous studies have extensively examined and reported on their presence in water and soil, less focus has been directed towards their occurrence in the air. This is chiefly due to the fact that pesticides generally exist in the air as trace-level pollutants, ranging from picograms per cubic meter (pg m
−3) to nanograms per cubic meter (ng m
−3). Pesticides can manifest in the air in diverse states, encompassing solids, gases, and liquids
[16][17]. Agricultural spraying activities, for instance, contribute to around 30–50% dispersion of most pesticides into the air
[17][18], primarily facilitated through drift (wind-carried) and evaporation. Subsequently, pesticides undergo volatilization from soil and plants, degradation, and photolysis, eventually integrating into the atmospheric environment. Airborne pesticides present a dual concern, as they not only pose immediate health risks through inhalation but can also settle on surfaces, leading to prolonged indoor contamination. Some studies have consistently highlighted the prevalence of organochlorine and organophosphate pesticides as the prominent groups detected in indoor air.
Studies have examined the temporal analysis of pesticide residues to pinpoint factors influencing the transportation and redistribution of these compounds within indoor environments. For example, Obendorf et al.
[18][40] discovered that the occurrence and quantity of pesticide residues, particularly chlorpyrifos, were elevated during the summer months—a correlation attributed to agricultural and horticultural practices. Conversely, greater quantities of insecticides like mecoprop, resmethrin, and tetramethrin were detected on flat surfaces during the winter, suggesting household application and potential redistribution within the indoor environment. In addition, a study conducted by Berger-Preiss and Elflein
[19][41] observed the presence of pesticides (such as pyrethroids, pyrethrum, and the synergist piperonyl butoxide) in an experimental house over a two-year period. The concentrations of these pesticides were the highest at the top of the room and gradually decreased towards the middle and lower areas
[20][42]. While some pesticides may degrade quickly through photodecomposition, others like DDT, chlordane, heptachlor, methoxychlor, dieldrin, and pentachlorophenol tend to persist indoors
[21][43].
Household pesticides are also found in dust and PM. The arena of dust and suspended particles becomes another tableau for household pesticides, as Rudel et al.
[21][43] revealed the prevalence of permethrins and the synergist piperonyl butoxide, woven into concentrations ranging from 1.7 to 17 µg g
−1. Wang et al.
[22][44] gathered dust samples from floors within residential homes and office spaces. Among the array of pesticides, it was observed that hexachlorobenzene’s presence in indoor settings could be attributed to its transfer from outdoor sources
[23][45]. On the other hand, chlordanes were linked to historical household usage
[24][46], while pyrethroids showed an association with ongoing household applications
[25][47].
2.2. Outdoor Air
Various pesticide groups, including organochlorine insecticides, and organophosphate insecticides, herbicides, and fungicides, have been detected in outdoor air samples from several countries such as France, Spain, China, Pakistan, Malaysia, South Africa, and the United States. Mirroring the trend observed with indoor pesticides, the organochlorine insecticides that have consistently surfaced as the most frequently detected compounds over the past 15 years include 4,4′-DDE, 4,4′-DDD, and endosulfan. Their concentrations ranged from 0.002 to 25.6 ng m
−3, 0.011 to 154 ng m
−3, and 0.0001 to 81.3 ng m
−3, respectively. Much like indoor environments, outdoor air contains traces of organophosphate insecticides such as chlorpyrifos, malathion, and diazinon, although typically at lower concentrations. However, their concentrations are commonly higher in indoor air compared to outdoor air due to their prevalent use for indoor pest control. Factors such as their proximity to indoor application sources, limited dispersion indoors, volatility, persistence, surface deposition, and potential re-suspension contribute to their higher concentrations. Outdoor air, in contrast, experiences lower exposure to these compounds due to differences in usage patterns and dispersion dynamics.
3. Advancements in Pesticide Sampling Techniques
3.1. Active Sampling Applications and Limitations
Active air sampling (AAS) is a technique that involves using a pump to collect gases, vapors, and particulates in a tube equipped with a sorbent bed or a size-selective sampler with a filter
[26][27][51,52] (
Figure 1). The volume of air sampled can be accurately measured using a flow meter. AAS is commonly employed for short-duration sampling, ranging from hours to days, making it suitable for monitoring daily or weekly variations
[28][53]. However, AAS can also be utilized for long-term sampling, requiring hundreds or thousands of samples to obtain annual data. For instance, Hung et al.
[29][54] utilized high-volume air samplers with a glass fiber filter and polyurethane foam (PUF) to monitor persistent organic pollutants (POPs) such as organochlorine pesticides and dichlorodiphenyltrichloroethanes in the Arctic region from 1993 to 2012. High-volume samplers effectively capture a higher amount of airborne pesticides compared to low-volume samplers
[30][55]. However, the use of high-volume sampling techniques may introduce some minor errors, such as gaseous compounds being adsorbed on deposited particles or filters (blow-on), while volatile compounds may desorb from the filter (blow-off)
[31][56]. To obtain a sufficient volume of pesticides in the air, Yusà et al.
[30][55] suggested considering the sampling objective, sampler flow rate, and the analytical method’s limit of detection. In addition to high-volume samplers, diffusion denuder systems have been proposed for measuring semi-volatile organic compounds (SVOCs) such as polycyclic aromatic hydrocarbons (PAHs), polychlorinated biphenyls (PCBs), organochlorine pesticides (OCPs), or carbonyl compounds in the atmosphere
[32][33][57,58].
Figure 1.
Diagram illustrating the active air sampling technique.
In terms of the filters employed in active air sampling, various materials have been used. Glass fiber filters
[34][63] and quartz fiber filters
[35][64] have been commonly utilized. According to Yusà et al.
[30][55], the diameter of these filters typically ranges from 9 to 30 cm, depending on the specific sampling technique.
When it comes to adsorbents for capturing gases, several materials have been employed. Polymeric phases such as polytetrafluoroethylene membranes (PTFE)
[13], Tenax TA
[36][65], XAD
®-2 resins
[37][66], XAD-4
[38][67], and PUF
[38][67] have been used. Among these, XAD
®-2 resins, which are hydrophobic copolymers of styrene-divinylbenzene resin, are widely preferred due to their high efficiency in trapping various types of herbicides, fungicides, and insecticides. PUF has also been extensively applied for monitoring OCPs and other pollutants such as PCBs, polybrominated diphenyl ethers (PBDEs), PAHs, and polychlorinated dibenzo-para-dioxins (PCDDs)
[30][39][40][41][55,68,69,70].
3.2. Passive Sampling Applications and Limitations
Passive air sampling (PAS) is a technique that relies on the natural diffusion of gaseous pesticides through adsorbents, eliminating the need for a pump
[42][73]. PAS typically operates at rates below 5 m
3 day
−1 [28][53]. A passive air sampler consists of a commercial accumulating medium with a high retention capacity for the target analytes (
Figure 2). The adoption of PAS has gained global popularity over active air sampling (AAS) due to several reasons. Firstly, PAS offers convenience for integrative sampling in remote locations, as it does not require electricity
[43][44][74,75]. Secondly, it allows for longer sampling durations, ranging from weeks to months, enabling the representation of monthly to yearly average exposure while minimizing spikes from episodic incidents
[28][30][43][53,55,74]. Thirdly, PAS exhibits a broader range for sampling different types of pesticides, including both persistent and less persistent ones
[45][32].
Figure 2.
Diagram illustrating the passive air sampling technique.
PAS generally employs various sorbent polymeric phases to capture pesticides. These include Tenax TA
®, PUF disks, sorbent-impregnated PUF (SIP), semi-permeable membrane devices, and carbon-based foams
[27][45][46][47][48][49][50][32,52,77,78,79,80,81]. PUF has been widely used since 2002, particularly for monitoring persistent compounds, but it has also been effective for sampling less persistent pesticides
[45][32]. In response to the limitation of PUF in detecting glyphosate, TIEM Integrated Environmental Engineering from Germany has developed a polyester filter that effectively captures both glyphosate and aminomethylphosphonic acid in ambient air
[51][52][38,82].
The choice of sorbent and housing design significantly affects the mass and type of collected particles. PUF and SIP can collect a representative portion of the particle phase, while XAD resins are suited for the gas phase
[53][83]. PUF-based PAS has been widely utilized due to its high capacity for detecting atmospheric pollutants, simplicity in structure, cost-effectiveness, and ease of operation
[48][54][79,84].
PUF disks have been employed to collect various airborne pesticides, including POPs and SVOCs such as OCPs and OPs
[55][56][57][58][59][60][86,87,88,89,90,91]. Nonetheless, Hayward et al.
[61][92] noted that while the PUF-PAS might have reached equilibrium with the atmosphere during deployment, the average air concentrations over extended periods did not significantly differ from those determined by AAS. They also suggested that for evaluating long-term air concentration trends in a cost-effective manner, utilizing fewer samples, the preferred approach would involve year-long XAD-PAS deployments. Moreover, PUF’s low effective surface area and limited reusability pose challenges, especially when using PLE at high temperatures
[27][52]. Despite the advantages of PAS, accurately measuring the specific volume of air passing through the adsorbent during exposure remains a drawback. To address this issue, Lévy et al.
[27][52] recommended calculating the sampling rate based on field data.
3.3. Evaluation of Sampling Techniques for Measuring Airborne Pesticides
The assessment of each factor’s impact was conducted on a scale from 1 to 4, with 1 denoting a low impact, 2 indicating a moderate low impact, 3 suggesting a moderate high impact, and 4 representing a high impact. As known, active sampling methods involve deliberately directing airflow through a sampler, leading to swift sample collection. In contrast, passive sampling relies on natural diffusion, which extends the sampling duration. Nonetheless, it is important to consider that this extended exposure can cause degradation, affecting both the chemical compounds being measured and the sorbent material itself. As a result, compound-specific degradation may introduce variability in sampler calibrations
[62][60]. Active sampling often achieves higher precision due to controlled airflow and consistent sample collection, while the precision of passive sampling is influenced by environmental factors like wind speed, temperature, and air concentrations. Various refinements in PAS have been introduced to address and control these factors, including the calibration of sampling rates, design modifications in sampler housing, and the use of performance reference compounds. However, despite these efforts, environmental variables can still introduce bias and errors in estimated air concentrations
[62][63][64][60,99,100].
4. Emerging Trends in Pesticide Extraction Techniques: Enhancing Efficiency and Analytical Performance
4.1. Pesticide Extraction Techniques
The extraction step plays a crucial role in separating pesticide residues after sampling. The most commonly used technique for extracting target analytes from the samplers is liquid-solid extraction, with Soxhlet extraction (SE) being the prevailing method
[65][101]. SE involves the use of a single solvent such as acetone or dichloromethane, or binary solvents like hexane-dichloromethane, dichloromethane-light petroleum, cyclohexane-acetone, or hexane-acetone
[30][55]. However, traditional SE has certain drawbacks, including its time-consuming nature and the potential for environmental harm due to solvent release
[66][102].
To address the concerns associated with SE and promote more environmentally friendly extraction techniques, alternative methods have been developed. These include liquid-solid extraction (LSE)
[13], PLE
[30][67][55,103], UAE
[68][104], and MAE
[69][105]. These techniques offer advantages over traditional SE, such as reduced extraction time, decreased solvent consumption, and enhanced extraction efficiency. LSE, PLE, UAE, and MAE have emerged as more environmentally friendly alternatives, contributing to the advancement of pesticide residue extraction methods. These approaches provide researchers with options that not only are efficient but also mitigate the potential negative impacts associated with traditional SE.
LSE is a commonly used technique for extracting pesticide residues but is often time-consuming, solvent-intensive, and laborious
[70][106]. In contrast, alternative extraction methods offer advantages such as rapidity, automation, selectivity, and reduced solvent consumption, making them more environmentally friendly options
[70][106]. Among these alternatives, PLE, also known as accelerated solvent extraction (ASE), has gained significant attention for the extraction of airborne pesticides due to its short extraction time, low solvent consumption, high contaminant yield, improved selectivity, and user-friendly system
[71][72][107,108]. PLE involves using an extracting solvent to flush a solid or semi-solid sample under intense heat (50–200 °C) and high pressures (500–3000 psi) for a short duration (around 10 min)
[30][73][55,109]. The efficiency of PLE depends on critical variables such as solvent selection, pressure, temperature, flush volume, extraction time, sorbent type, and the number of extraction cycles, necessitating optimization procedures
[71][107].
UAE, also known as sonication, is an environmentally friendly technique used as an alternative method for extracting particle pollution. This method utilizes acoustic waves to generate cavitation bubbles, which enhance the solubility of analytes and the diffusion of solvents within the matrix
[68][104]. UAE offers several advantages, including a significant reduction in extraction time, lower solvent usage, fewer opportunities for contamination and analyte losses, and the development of eco-friendly and cost-effective methods with increased productivity
[73][74][75][76][109,113,114,115].
It should be noted that UAE may be less precise than automated methods like PLE or MAE, especially when applied to matrices with strong interferences
[77][117]. Furthermore, the sonication involved in UAE has the potential to damage sampling filters, which could result in the release of particles during the extraction process
[73][109]. These factors should be taken into consideration when selecting the appropriate extraction method for airborne pesticide analysis. In summary, UAE offers a greener alternative for extracting pesticides from PM, providing advantages such as reduced extraction time, decreased solvent usage, and enhanced productivity. Despite its limitations in certain matrix types and the potential for filter damage, UAE has demonstrated successful applications in extracting airborne pesticides from solids and PM.
4.2. Analytical Performance of Pesticide Extraction Methods
When conducting a comparative assessment of various extraction methods for analyzing airborne pesticides through the utilization of a spider chart (
Figure 34), several crucial factors emerge. For instance, the operational time varies, with SE entailing time-consuming processes due to prolonged extraction cycles. In contrast, methods like PLE and MAE expedite procedures by capitalizing on elevated temperature and pressure or microwave heating. Extraction efficiency also displays variability: SE and PLE excel in efficiency, particularly for heat-sensitive compounds in the case of PLE, while UAE and MAE ensure effectiveness through improved mass transfer or swift and uniform heating. The spectrum of operational simplicity ranges from complexity with SE to relative simplicity with UAE and MAE.
Figure 34. Comparative evaluation of extraction methods for airborne pesticide analysis. Level of impact: 1 = Low impact; 2 = Middle low impact; 3 = Middle high impact; 4 = High impact.
5. Advancements in Pesticide Analytical Methods
5.1. Advances in GC for Pesticide Detection in Air
The measurement of atmospheric pesticides poses a significant challenge, particularly when dealing with concentrations below 2 ng/m
3 [43][74]. In both AAS and PAS, an extraction process is necessary to release the trapped pesticides from the media. Two common techniques for extraction are organic solvent desorption and thermal desorption (TD). The organic solvent method involves multiple extraction and concentration steps, resulting in higher uncertainty and a more time-consuming process
[43][74]. In contrast, TD eliminates the need for a concentration step, and it can be coupled with GC-MS to achieve lower quantification limits compared to organic solvent extraction. GC is widely utilized for pesticide measurements in combination with various element-specific detectors, such as atomic emission (AED), electron capture (ECD), sulfur chemiluminescence (SCD), nitrogen-phosphorus detection (NPD), and MS.
GC-ECD is commonly used for the analysis of OCPs in food and environmental samples due to its high separation efficiency, sensitivity, and cost-effectiveness
[78][79][80][123,124,125]. Non-polar or semi-polar columns with dimensions of 30 m × 0.32 mm ID × 0.25 mm film thickness are typically employed, and helium is commonly used as the carrier gas
[30][55]. However, it is important to note that GC-ECD is particularly well suited for the analysis of thermally stable compounds due to its thermal desorption capabilities. Given this consideration, Liu et al.
[80][125] recommended using GC-negative chemical ionization-mass spectrometry (GC-NCI-MS) instead of GC-ECD for analyzing OCPs and hexabromobiphenyls (HBBs) in atmospheric PM and soil samples. They found that GC-NCI-MS provided better sensitivity and robustness compared to GC-ECD.
Nowadays, GC-MS is the predominant technique used for determining pesticides in the air, primarily due to the volatility of most compounds
[81][127]. Capillary columns with various trade names, such as MDN-5, DB-5, TR-5MS, SGE-BPX5, or V5-MS, are commonly employed
[9][15][82][83][84][9,15,128,129,130]. Helium gas with a purity of 99.99% or argon C50 gas with a purity of 99.99% is typically used as the carrier gas
[82][128]. In GC-MS analysis, the major application for the MS analyzer is the use of quadrupole in selected ion monitoring (SIM) mode, which offers higher sensitivity compared to the full scan mode
[30][55]. The MS analyzer is primarily operated in the electron ionization (EI) mode, especially in multi-residue analysis.
Pesticide quantification is based on the GC-MS peak area, and external calibration curves are generated by directly injecting analytical standards of the target pesticides. However, the use of tandem mass spectrometry (MS/MS) has recently been proposed as a valuable tool due to its increased selectivity and reduced mass spectral noise
[85][131]. Lee and Jo
[86][132] and Wu
[87][133] emphasized the excellent selectivity and sensitivity provided by GC-MS/MS with a triple quadrupole (QqQ) analyzer. It can be operated in multiple reaction monitoring (SRM) mode, enabling more reliable identification and quantification of target analytes.
To ensure satisfactory sensitivity for each pesticide, the precise optimization of MS/MS variables is required. The initial step involves selecting the parent ions from each pesticide using a full scanning spectra mode. Subsequently, the precursor ions are accumulated and isolated in the ion trap, followed by fragmentation through collision-induced dissociation
[30][55]. Among the resulting product ions for each congener, the two most prominent ones are selected. The optimization process can be carried out using either the approach of changing one factor at a time or the design of experiment procedure
[88][120]. This optimization ensures that the MS/MS settings are fine-tuned for optimal sensitivity and the accurate identification of target pesticides.
5.2. Advances in LC for Pesticide Detection in Air
GC-MS in SIM mode and GC-MS/MS-QqQ are commonly used for GC-amenable pesticides. However, for non-GC amenable pesticides, such as polar, non-volatile, and thermolabile compounds like herbicides, carbamates, triazines, phenoxy acids, or neonicotinoids, HPLC has been proposed as an alternative method. HPLC is suitable for the separation and quantification of these pesticides, especially when they require derivatization to enhance volatility, thermal stability, and sensitivity, or to address the limitations of GC-MS methods. Despite the prevalence of GC-MS techniques, there are limited studies that have applied HPLC methods for pesticide analysis in ambient air.
5.3. Other Advances in Pesticide Detection Methods
IMS has recently gained significant prominence as a robust separation technique owing to its unique design, high sensitivity, rapid response time, operation at ambient pressure, and capability to effectively separate isomeric compounds. This versatile technique finds applications in diverse fields such as chemical weapons detection, explosives analysis, pharmaceutical screening, and environmental monitoring
[89][143]. In IMS, gas-phase ions are generated by ionizing neutral molecules, and subsequently, they are separated based on their distinct velocities within an electric field before quantification. The separation efficiency is influenced by factors such as temperature, pressure, and the molecular properties of the drift gas.
Attenuated total reflectance–Fourier-transform infrared spectroscopy (ATR–FTIR) spectroscopy is a widely utilized technique for the chemical characterization of environmental samples
[90][91][144,145]. ATR, as a sampling mode, enhances the FTIR signal obtained from the surfaces of samples, making it a promising tool for detecting organic pesticides and hazardous mineral compounds such as asbestos, which is a re-emerging contaminant in environmental matrices
[14]. While ATR-FTIR offers rapid and non-invasive analysis, with improved reproducibility compared to traditional analytical methods, it has limitations in the molecular-level identification of contaminants of emerging concern. To address this challenge, a complementary spectroscopic analysis data of non-target analytes is recommended.
Synchrotron radiation–attenuated total reflectance–Fourier-transform infrared spectroscopy (SR-ATR-FTIR) represents a sophisticated analytical approach leveraging synchrotron light’s unique qualities
[92][93][147,148]. Unlike traditional FTIR, it offers brightness, spatial precision, and energy versatility, with distinct advantages for airborne pesticide analysis in environmental samples. This technique requires the preliminary processing or extraction of environmental components like soil, water, or plant tissue. With SR-FTIR, the interaction between emitted infrared radiation and the sample generates molecular fingerprints. Exposure to potent synchrotron radiation enables the selective absorption of IR light by pesticide functional groups, yielding specific spectral peaks. The exceptional resolution capability of SR-ATR-FTIR facilitates microscale exploration, uncovering pesticide traces undetected by conventional methods like GC-MS and LC-MS. Spectral examination aids in identifying pesticide-specific absorption peaks, enabling the qualitative and quantitative assessment of residues using reference spectra. Overall, SR-ATR-FTIR stands as a powerful tool for intricate airborne pesticide analysis within complex environmental settings.
5.4. Analytical Performance of Pesticide Detection Techniques
When employing a spider chart to assess various techniques for measuring airborne pesticides across different factors (
Figure 45), several pivotal considerations become apparent. In terms of operational time, GC techniques exhibit variations depending on the specific method utilized (e.g., GC-MS, GC-FID). This encompasses stages such as sample preparation, injection, column separation, and detection, collectively contributing to a distinct timeframe. Similarly, the HPLC method requires several sequential processes, making it relatively time-consuming. In contrast, IMS expedites analyses due to its rapid ion mobility separation process. ATR–FTIR and SR-ATR-FTIR offer quicker analysis due to minimal sample preparation and the efficiency of FTIR techniques. Likewise, leveraging synchrotron sources for rapid scanning enables time-resolved studies, facilitating insights into pesticide degradation and transformation in the environment. HPAEC involves multiple steps similar to GC and HPLC, contributing to a moderately paced operational timeframe.
Figure 45. Comparative assessment of analytical techniques for airborne pesticide analysis. Level of impact: 1 = Low impact; 2 = Middle low impact; 3 = Middle high impact; 4 = High impact.
Regarding sensitivity, both GC and HPLC methodologies are renowned for their exceptional sensitivity, enabling the detection of trace compounds, including airborne pesticides, within complex matrices. Similarly, IMS also demonstrates sensitivity to specific compound classes, although it might not reach the same level as chromatography-based methods. ATR–FTIR recognizes functional groups and chemical bonds, albeit sensitivity toward specific pesticides may exhibit variance. HPAEC’s sensitivity extends to anions, encompassing selective pesticides and aligning with specific applications. SR-ATR-FTIR offers high sensitivity, making it suitable for detecting trace amounts of pesticide residues.
In assessing operational simplicity, GC and HPLC demand adept analysts and intricate hardware, diminishing their operational ease. IMS stands out as comparatively user-friendly, necessitating minimal analyst expertise when compared to chromatographic methods. ATR–FTIR demonstrates relative straightforwardness with minimal sample preparation and swift analysis. HPAEC introduces a degree of complexity, manageable with proper training. Meanwhile, SR-ATR-FTIR demands a certain level of skill but offers a harmonious blend of advanced capabilities and user-friendliness.
When evaluating the requisite analyst skill, GC and HPLC demand proficient analysts for meticulous method development, sample preparation, and data interpretation. IMS calls for less analyst expertise in comparison to intricate chromatographic approaches. ATR–FTIR and SR-ATR-FTIR mandate basic training for operational competence and spectral interpretation. HPAEC, on the other hand, necessitates adept analysts, proficient in method development, sample preparation, and data interpretation.
In terms of cost analysis, GC and HPLC instruments, along with their consumables and maintenance, incur moderate expenses. IMS instruments are often affordable, with generally lower operational costs in comparison to chromatography. ATR–FTIR instruments fall within a moderate cost range, with routine analyses proving cost-effective. For HPAEC, instruments and consumables hold moderate costs, justified by the technique’s specificity. On the other hand, SR-ATR-FTIR entails higher costs due to synchrotron access and equipment requirements, but its capabilities substantiate the investment.
From an environmental sustainability standpoint, GC and HPLC utilize solvents that can potentially impact the environment. However, ongoing efforts are directed towards adopting greener practices to mitigate these effects. In contrast, IMS stands out for its reduced solvent usage and minimized waste generation, which contribute to a more environmentally sustainable approach. Both ATR–FTIR and SR-ATR-FTIR require minimal reagents and generate limited waste, further reinforcing their positive environmental profile.
Turning to health implications, GC and HPLC demand the vigilant handling of solvents and analytes to mitigate potential health risks associated with hazardous substances. On the other hand, IMS involves fewer harmful chemicals, alleviating health concerns for analysts. The ATR–FTIR and SR-ATR-FTIR methods entail minimal exposure to hazardous chemicals, which significantly reduces health risks for practitioners. HPAEC necessitates the proper management of solvents, reagents, and analytes to effectively minimize potential health hazards. This comprehensive evaluation highlights the significance of considering environmental sustainability and health aspects when choosing among these analytical techniques.