Microfluidic-Based Oxygen Sensors: Comparison
Please note this is a comparison between Version 1 by Patricia Khashayar and Version 3 by Vicky Zhou.

Oxygen (O2) quantification is essential for assessing cell metabolism, and its consumption in cell culture is an important indicator of cell viability. Recent advances in microfluidics have made O2 sensing a crucial feature for organ-on-chip (OOC) devices for various biomedical applications. OOC O2 sensors can be categorized, based on their transducer type, into two main groups, optical and electrochemical.

  • oxygen sensors
  • microfluidics
  • organ-on-chips (OOCs)
  • on-chip monitoring

1. Introduction

Oxygen (O2) is one of the main components of cellular respiration and energy production [1]. The availability of O2 is a key metric that defines the pathway of adenosine triphosphate (ATP) generation and its resultant metabolites that serve as the living cell’s energy source [2][3][2,3]. In a high O2 environment, ATP is synthesized by the phosphorylation of the precursor molecule adenosine diphosphate (ADP). This process, thus called aerobic respiration, requires an adequate level of O2. It consists of the coupling of electron transport and oxidative phosphorylation, where O2 acts as the final electron acceptor from the oxidation of glucose and/or glycogen [4]. In low O2 environments, conversely, ATP is generated at an inefficient but rapid rate via a process called anaerobic glycolysis, where glucose and glycogen are metabolized to pyruvate and lactate in the absence of O2. This pathway is important in the functions of vital organs such as the kidney and retina as well as in tumor formation [5]. O2 availability determines the metabolic pathway that generates energy for cell function and survival and therefore is significantly important to measure for bioassays, cell culture and diagnostic applications.
Precise control of a small amount of fluid is possible inside microfabricated channels of microfluidic technology [6]. Integrated microfluidic chips are capable of performing highly sensitive and low-cost analyses. These platforms can be integrated with new technologies with cell culture/organoid studies at high temporal and spatial resolution. For instance, microfluidic platforms can quantitatively monitor cellular signals and cell secretions using well-developed cell-culture methods on microchips [7]. O2 measurement can be performed using sensors integrated into microfluidic chips for monitoring the metabolism and viability of cell, tissue, and organ [8][9][8,9]. Moreover, novel sensors embedded inside implantable microchips have recently been used for real-time in vivo oximetry in human or animal bodies [10]. Sensors can also be used to detect and quantify various analytes in a complex biological environment [11][12][13][14][15][16][11,12,13,14,15,16]. Therefore, a combination of integrated sensors is required for online and non-invasive monitoring of the cell intake, secreted metabolites, and microenvironment status for on-chip microfluidic studies such as OOC and bioreactors [12][13][12,13].
On-chip monitoring of O2 is pivotal in OOCs. The low concentration of O2 inside a small chip and its crucial biological role in cell metabolism and function would emphasize the need for its precise and selective quantification in the confined environment of a microfluidic channel or chamber [9][17][9,17]. With recent advances in microfluidics-based cell and tissue studies, such as OOC technologies, various sensors have been integrated into chips to monitor the microphysiological parameters of cells [12][18][19][20][21][22][23][12,18,19,20,21,22,23]. OOCs have the potential to better mimic human organs compared to the traditional in vitro models, and thus they can reduce the need for animal models in interventions such as the studies on drug efficacy and toxicity [24][25][26][24,25,26]. To model the function of many organs, such as the pancreas [27], brain [28][29][30][28,29,30], liver [31], vascular system [32], Gut [33], multiorgan approaches [34] and body-on-the-chip, OOC-integrated O2 sensing techniques have been used to culture cell monolayers, three dimensional (3D) cultures, spheroids, organoids and stem cells. They have also been used to model tumor microenvironment by mimicking the extracellular matrix (ECM) [35] and 3D culture of cancer cells [36][37][36,37].
Electrochemical and optical sensors are the main transducers for on-chip O2 monitoring [38], as they are precise, selective, and easy to miniaturize and implement inside chips [18][39][18,39]. In addition, micro and nanomaterials along with the innovative designs and polymers, and commercial readout devices are used for signal amplification to overcome the limitation related to the measurement of low O2 levels with electrochemical and optical sensors [13][40][41][13,40,41]. Here, we review the recent innovations in O2 sensors integrated into microfluidic chips, including OOC devices. We have categorized them based on their transducer type into two main sections, namely optical and electrochemical. We also discuss recent innovations and their advantages and disadvantages. Finally, we provide a comprehensive discussion of the current advances in the design, fabrication and application of optical and electrochemical O2 sensors.
Previous reviews have covered related topics such as oxygen control (2016) [9], general optical imaging and sensing (2014) [39], microfluidic OOC sensors (2010) [40], and other microphysiological sensors of OOCs [12][13][18][38][12,13,18,38]; however, there are no recent publications critically discussing the current advancements in OOC O2 sensing.

2. Oxygen Sensors in On-Chip Systems

Methods of on-chip O2 measurement can be categorized into two groups of sensors: (i) optical and (ii) electrochemical. Here, we highlight the application of these two types of sensors in on-chip studies, explain their mechanism of action, and discuss their advantages and disadvantages. Table 1 represents the summary of advantages and disadvantages of optical and electrochemical methods for on-chip O2 measurement.
Table 1.
Comparison of chip-based electrochemical and optical O
2
sensors.
Method Advantages Limitations
Optical

2.1. Optical Methods

Table 2 represents the summary of most recent on-chip optical O2 sensors and their features, applied dyes, and their advantages.
Table 2.
Characteristics of chip-based optical O
2
sensors.
Optical O2 Sensor Application Dye Advantages References
Polystyrene chip, pore network structure, used solvent-induced fluorophore impregnation (SIFI) method for dye layer
  • Precise, sensitive and selective
  • Easy to miniaturize
  • Nano and microparticles for dye protecting
  • Non-invasive, and contact-free
  • Easy to use and operate
  • Commercial dyes and read-out devices
  • Simple handling and sterilizing of the chip
  • Less need to recalibrate
  • Multiplex measurement in different chip spots
  • Simultaneously measure pH and metabolites
  • Complicated integration into the chip
  • Possible dyes bleaching
  • Sometimes needs microscope
Cell Platinum(II) tetrakis(pentafluorophenyl)porphyrin (PtTFPP)PtTFPP Enhanced sensitivity and stability, non-invasive, can be used for gas and dissolved O2 [41]
PDMS-container structure, and the glass substrate 105 cells/mL Short response time (6.9 s) [52][83]
Electrochemical
  • Precise, sensitive and selective
  • Easily miniaturized/implemented inside chips
  • Nano and microparticles for dye protecting
  • Possible use of commercial read-out devices
  • Short response time
  • High sensitivity
  • Label-free
  • Several electrodes
  • Several surface modifications
  • Several designs and polymers can be used easily
Poly (dimethylsiloxane) (PDMS) chip with glass layer coverage, applied oxygen gradientPDMS chip with glass layer coverage, applied oxygen gradient
  • Invasive and consume oxygen
  • Expensive integration in the chip
  • Requires special instrument and skilled operators
Liver Pt(II) Octaethylporphine (PtOEP)PtOEP Wide dynamic range, continuous measurement, non-invasive, worked in different flow rates [42]
Low-temperature co-fired ceramic (LTCC) in an improved Clark-type DO sensor Up to 8.1 mg/L easy fabrication, flexible configuration, short response time (10.9 s), real-time detection [53][84] Cyclic olefin copolymer-based chip Lung PtTPTBPF Simultaneous O
pHEMA hydrogel layer with electrolyte and PDMS as gas-permeable membrane 0.121 μA cm−2 μM−12 and pH, stop/flow measurements, long term stability (10 days), non-invasive zero analyte consumption, 1-point calibration, long-term stability [54][85[43]
] PDMS chip, applied oxygen gradient Cancer PtOEPK
PPy as the internal contact layer between polymeric sensitive membrane and gold 0.11 ± 0.02 mg L−1Photostable, reusable, non-invasive Low cost, good performance and long-term potential stability [55][86][44]
PDMS chip, silica microparticles Cancer Ru(dpp)
Multi-sensor glass-chip with a PDMS imprinted microfluidic channel gridSimple fabrication and handling, real-time, spatially-resolved measurements, low photobleaching, High sensitivity 100 pA per each 1% O2 Transparent for microscopic observation, cheap, high sensitivity [56][87][45]
Poly (methyl methacrylate) (PMMA) chip, polystyrene microspheresPMMA chip, polystyrene microspheres Embryo study Pt-porphyrin
Biocompatible glass chip fabricated using a hybrid thin film and laminate technologies 0.735 μA μM−1 cmSimultaneous O2 and pH, long-term measurement, highly sensitive for single embryo analysis −2 Low O2[46]
consumption on the electrode, long-term stability [57][88] Poly (dimethylsiloxane) (PDMS) chip, polystyrene microbeadsPDMS chip, polystyrene microbeads Liver ruthenium-phenanthroline (RuP) Every 15 min for 28 days measurement, without a decrease in signal loss and toxicity, simultaneous glucose and lactate measurements
Biocompatible PDMS biochip with Au/Nafion electrodes 50 mmol L−1[47]
real-time and continuous O2 monitoring in dynamic flow conditions [58][89] Glass chip, nanoparticle probes Stem cell platinum(II) meso-bis(pentafluorophenyl)bis(4-bromophenyl)porphyrin (PtTFPPBr2)PtTFPPBr2
Kapton tape with embedded spirally rolled MicrochannelsHighly sensitive, real-time, label-free, high-intensity fluorescence emission, cell permeability 12.89 nA mmHg −1 O2 and temperature sensors, embedded spirally rolled microchannels [59][90][48] Teflon fluorinated ethylene propylene (FEP) tubing, poly(styrene-block-vinylpyrrolidone) nanobeads Bacteria Platinum (II)-meso-tetra(4-fluorophenyl)tetrabenzoporphyrin (PtTPTBPF)PtTPTBPF Minimized background fluorescence, simultaneous measurement, highly soluble and disperse nanobeads, prevents any interferences from biomolecules, short response times, no dye leaching, and long storage periods [49]
Silicon/glass chip, core−shell nanosensors (poly(styrene-blockvinylpyrrolidone) Fibroblast cell PtTPTBPF Simultaneous O2 and pH, contactless and inexpensive read-out, high ionic strength, highly stable, online monitoring [50]
Glass chip, polymeric nanoparticles Cell Pt(II) benzoporphyrin Highly stable at different pH, ultrafast response (less than 0.2 s), no leaching, repeatable [51]

2.2. Electrochemical Methods

Table 3 summarizes the components, advantages and LOD of the selected ElectrochemicalC (EC) sensors for the measurement of DO.
Table 3.
Characteristics of chip-based electrochemical O
2
sensors.
EC-Based O2 Sensor LOD Advantages References
ElecCell technological platform using PVD
6 pA/s
low-cost, easy to use and reproducible portable chip
[
60][91]
ultra-microelectrode array (UMEA) 0.49 nAs−0.5/mg/L Ultra-short response time (<5 ms), 10 times lower O2 consumption [61][92]
Multi-planar SPE sensor coupled with cultivation cell wells 3 mg/L Continuous long-term O2 measurement, sensor reutilization [62][93]
Inkjet printing (IJP) DO sensors on the delicate porous substrate 28 ± 1 nA L mg−1 low O2 consumption on electrodes, short response time (60 s) [63][94]
Electrochemical microsensors combined with spheroid technology NM fast, precise, and continuous long-term measurement of metabolic directly in the microwell [64][95]
Spheroid on chip NM Real-time monitoring of metabolic activity and automated assays for toxicity evaluation [65][96]
NM: not mentioned.

3. Conclusions and Future Perspectives

Herein discussed recent advancements of oxygen sensors in on-chip systems and categorized them in two main groups: optical and electrochemical methods. The optical methods are reported to be more sensitive, easier to operate and cheaper compared to the EC methods. In most cases, they do not consume O2 during the process. In addition, they are compatible with commercially available luminescent dyes and optical readout devices, even fluorescent microscope which is convenient and available in most cell and tissue process centers. These sensors can be used for contactless monitoring by adding a sensing spot outside the chip readout or optical fiber, making the handling and sterilizing of the chip simpler. In addition, they do not need recalibration or experience decay over time. Compared with EC methods, therefore, these techniques are more commonly used in OOC applications. This is especially important for low concentrations of the sample when the stability and reusability of the sensor and its remaining intact are crucial. On the other hand, the EC methods have a shorter response time, and in most cases, a higher sensitivity than the previous group. In most cases, they can be used as label-free sensors, which again reduces the cost of sensor fabrication compared with optical ones. However, their integration in the chip is relatively expensive, complicated, and sometimes requires special relatively expensive instruments (Potentiostat/Galvanostat) and skilled operators.
In the future, the use of novel materials, fabrication techniques, and chemical/physical surface modifications can help facilitate the fabrication steps, reducing the price and required specialty, making the chip-integrated O2 sensors more affordable. Although there are few examples of micro- and nanomaterials used for the chip-based O2 sensors of either type, the field is rapidly progressing. It is expected that a variety of materials with various properties to help with O2 sensing will be used for these sensors in the near future to improve their sensitivity. This is mainly because these materials are believed to increase the active surface area, enhance the stability of the surface and increase the glow of the dyes. For instance, graphene and quantum dots can be considered as potential materials due to their exceptional optical and electrochemical properties. Micro- and nanostructures of novel metal oxides are other examples of such materials because of their catalytic activity. In addition, quantum dots (QDs) are photostable and their excitement spectra are broad, while their emission spectra are size-tunable, which makes them suitable for optical sensors. Further, using a smartphone to quantify the emitted signal can significantly simplify the measurement process. Artificial intelligence and innovations in image and signal processing help enhance the sensitivity and specificity of O2 sensors. Additionally, using 3D printing to manufacture the sensors can result in flexible, rapid and low-cost designs and integration of sensors in OOCs, and improve the sensitivity of the O2 sensors inside the chips. Three-dimensional printing techniques can also be used to build customizable optical holders and modules that can be integrated with mobile phones or other portable detectors.