Biopotential Signal Monitoring Systems in Rehabilitation: Comparison
Please note this is a comparison between Version 2 by Nora Tang and Version 1 by Patrizia Vizza.

Rehabilitation consists of an iterative process involving assessments and specialized training, which unfortunately are often limited by healthcare centres’ restricted resources. To overcome this limitation, wearable technology should be an important, potential and valid solution to objectively assess and monitor patients inside and/or outside clinical environments. The information extracted by the use of this technology should provide a more detailed evaluation of the impairment, also allowing the identification of rehabilitation therapies.

  • biomedical signal
  • monitoring system
  • rehabilitation
  • signal processing

1. Introduction

Biomedical wearable sensors allow the measurement of physiologic parameters in a continuous, real-time and non-invasive way, including a wide range of advances in electrocardiogram (ECG), electromyogram (EMG) and electroencephalogram (EEG)-based sensing platforms [1,2,3]. These platforms and their related sensors have different diagnostic and monitoring applications [4,5]. For example, physiological monitoring could support both diagnosis and ongoing treatment for many diseases involving movement disorders [6]. Furthermore, home-based motion sensing could assist the subject in rehabilitation path and falls prevention, helping him/her improve his/her independence and lifestyle [7]. Moreover, sensors acquire and analyze biomedical signals to monitoring the effectiveness of home-based rehabilitation therapies, for example, in stroke survivors, in patients undergoing surgery, in subjects involved in accidents or to evaluating the use of mobility assistive devices in older adults [8,9,10]. Moreover, the monitoring of physiological signals and parameters could also be a good support in many clinical and non-clinical applications, e.g., in sporting activities to evaluate performance and physical condition of athletes [1,11], in postural control to correct stability, or in a physiotherapy context after injury [12,13]. Generally, the measurement of physical activity parameters aids in guiding many types of applications; e.g., (i) monitoring physical activity during rehabilitation or in a physical therapy setting; (ii) evaluating the success of an intervention and tracking physical activity post-surgery; (iii) evaluating patient mobility; (iv) all (risk) detection; and (v) monitoring physical activity in patients with chronic diseases and disabilities involving movement disorders. Most recently, the coronavirus disease 2019 (COVID-19) pandemic has affected access to standard rehabilitation services, highlighting the need to define new rehabilitation perspectives as telemedicine [14,15]. During this period, the rehabilitation concept considerably changed: the need for home medical assistance for a new idea of rehabilitation phase in older people both with and without COVID-19, in patients affected by neuro-motor disease, in subjects with limited movements after injury or accident and athletes is becoming essential to help these people in maintaining their daily activities. In this context, tele-rehabilitation became an effective and well-accepted method of providing outpatient and community rehabilitation services to support family and caregivers in the assessment of the home environment, patient monitoring and outpatient therapies [16,17,18]. In emergencies such as the COVID 19 pandemic, access to health services is restricted due to the risk of infections and limitations of health resources [14,19]. For this reason, telemedicine services have proved extremely useful by providing home monitoring and rehabilitation solutions and thus minimizing the risk of infection. Survivors of COVID-19-associated pneumonia may experience a long-term reduction in functional capacity and muscle strength. Telerehabilitation (TR) programs could be effective for patients after COVID-19 [14]. However, few studies have assessed whether telerehabilitation for COVID-19 patients is an effective tool. In [20,21], telerehabilitation programs consist of home exercises for aerobic reconditioning, muscle strengthening, and healthy lifestyle education. The physiotherapist (PT) contacts the patient via video call via a dedicated platform to monitor progress. Moreover, physicians can add chest physiotherapy exercises for lung expansion and strengthening of the respiratory muscles. In these programs, a pulse oximeter as a monitoring device is also used.

Many healthcare devices for rehabilitation provide biosignals, such as blood pressure, blood glucose levels, EEGs, ECGs and EMGs [22]. The main bioelectrical signals are generated by the heart, the brain and the muscles, producing ECGs, EEGs and EMGs, respectively. ECG, EEG and EMG signals are characterized by low amplitude (generally, expressed in mV—millivolts) and low operating frequencies, from frequency Hz to some kHz range [23]. The acquisition, analysis and interpretation of these signals are fully reported in the literature [24,25,26,27,28,29,30,31]. Physical activity is often associated with the cardiovascular and muscular systems. Therefore, electrical signal variations cause ECG and EMG during athletic activities, and they are essential and commonly adopted parameters for healthcare management and rehabilitation protocols. In particular, EMG signal is the typical clinical recording method used to diagnose and monitor neuromuscular behaviours. Surface EMG (sEMG) allows extraction of information on muscle activation during a movement or effort, identifying impairment and functional alteration useful in clinical evaluation [32,33]. This information can be presented in different forms (e.g., amplitude, timing, morphology, muscle fibre conduction velocity or muscle coordination). They are relevant in many fields, from orthopaedics and neurorehabilitation to movement analysis in exercise and sport or aging [34,35]. This review aims to focus on EMG signal acquisition devices, also combined with other biosignals ECG and EEG, in rehabilitation pathways, especially for telemedicine applications. This contribution is proposed as a review by addressing questions such as (i) what are the most recent contributions in literature? (ii) what are the commonly used medical devices? (iii) how do these contributions and medical devices support physiological monitoring in rehabilitation? and (iv) what are the future directions and opportunities for EMG signal acquisition and analysis in a rehabiliation context? Many reviews are presented in the literature regarding biosignal acquisition devices for rehabilitation applications, but to the best of our knowledge, EMG signal has been considered only in specific context for single review. This review is thus a general but detailed comprehensive overview of EMG monitoring systems aiming to resume and to discuss the different and important solutions of EMG applications in different rehabilitation contexts. The key contributions of this review are outlined as follows. Section 2 presents a summary of basic concepts such as bioelectrical signalss and medical devices aiming to identify their main features and characteristics. These concepts are required background knowledge to this review. Section 3 reports the research methodology applied to define this review. Section 4 proposes an intense state-of-the-art about the last 5 years’ contributions in wearable sensors and platforms to summarize recent developments in the field of wearable sensors and systems used for rehabilitation. In Section 5 , most commercial biomedical signal acquisition and processing sensors and devices have been reported comparing physical and performance characteristics. Section 6 has also been formulated to summarize and discuss the literature on EMG signal acquisition devices in rehabilitation and to explores their challenges and their future direction. Finally, Section 7 provides conclusions of the article. 

2. Wearable Devices for Rehabilitation

Generally, these systems present heavy drawbacks regarding the limitation in acquiring and sending data at high rates, the low energy efficiency and the restricted portability due to their large size and weight. To overcome these limitations and make these systems more efficient, wearable devices are becoming essential in daily and clinical practice to allow continuous monitoring of human activity in terms of changes in biological signals. The increasing trends of wearable devices and the multimodal acquisition of different biosignals are crucial for advancing disease-diagnosis and treatment. Wearable devices perform activity monitoring through two main processes: (i) data acquisition and preprocessing; (ii) transmission, analysis and classification of acquired data. Signal preprocessing, for example, includes amplification and filtering stage; signal analysis, instead, involves averaging or extraction of relevant features to be used as training data for classifier [83].

In literature, many contributions are available concerning the design and the implementation of wearable sensors aiming to define platforms of multimodal acquisition and recognition of different biosignals, such as electroencephalography, electromyography and electrocardiography, for continuous and automatic monitoring of human health status, improving diagnosis, follow-up and therapeutic strategies of several disorders. Wearable devices usually involve smart sensors to detect and monitor a set of physiological parameters aiming to support their continuous monitoring for diagnostic, therapeutic and control purposes [84]. The great demand of the aging population for healthcare management needs the use of these wearable medical devices to monitor personal health information in real-time to prevent diseases and emergency health risks. Today, many wearable healthcare devices provide biosignals, such as EEGs, ECGs, EMGs, blood pressure or blood glucose levels. Electrical signal variations cause ECG and EMG during muscular activities, and they are important and commonly adopted parameters for healthcare management and rehabilitation protocols.

Liu et al. in [85] propose a portable and wireless acquisition system to acquire physiological signals. The system mainly consists of a portable device, a graphic user interface (GUI) and an application program for displaying the signals on a computer or a smart device. This device is characterized by eight measuring channels, a powerful microcontroller unit, a lithium battery, Bluetooth 3.0 data transmission and a built-in 2 GB flash memory. The results show that as this system can measure signals in real-time, supporting physicians and researchers can perform experiments collecting physiological signals of interest.

A summary of these contributions about wearable monitoring systems chosen among the papers in the literature published in the last 5 years is made in Table 21 .

Table 1. Summary of selected wearable monitoring system included in this review.
AuthorsSignalsChannelsPlatform CharacteristicsFeatures
Tran et al., 2021Bio-potentials4 channelsFour-channel neural recording analog front-end composed by a low-noise amplifier (LNA), a programmable gain amplifier (PGA) and buffers; 4-to-1 multiplexer (MUX) and analog-to-digital converter (ADC)Programmable gain from 45 dB to 63 dB, input-referred noise of 3.16 μVRMS within the 10 kHz bandwidth, noise efficiency factor of 2.04, power efficiency factor of 4.16, power consumption of 2.82 μW per channel powered from 1 V supply voltage
Yin et al., 2021Bio-potentials, impedance respirationSingle 1 channelOversampling and fast digital lock-in technology, ADS1294R, STM32F103RET6 for signal processingImprove the common-mode rejection ratio (CMRR) and the signal-to-noise ratio (SNR) of the signal
Zhao et al., 2020ECG/EMGN.A.Low-energy Bluetooth moduleWearable monitoring device, software platform for data analysis
Biagetti et al., 2020Bio-potentials3 channelsSix electrodes, 24 bits of resolution and a sampling rate up to 3.2 kHz for each channel, Bluetooth Low Energy wireless linkWireless sensor, real-time acquisition, maximization of the available bandwidth, reliability of the transmission
Nakamura et al., 2020ECG/EMGN.A.Analog front-end (AFE)Capacitive measurements
Liu et al., 2019Bio-potentials8 channelsPowerful microcontroller unit, lithium battery, Bluetooth 3.0 data transmission and built-in 2 GB flash memoryPortable device with a graphic user interface (GUI) and an application program for displaying the signals on a computer or a smart device
Park et al., 2018Bio-potentials128 channelsEnergy-efficient integrated circuit architecture of a Δ-modulated ΔΣ AFE with multi-shank neural probes connected to individual AFEsThe Δ-ΔΣ AFE is characterized by a consume of each single-channel AFE of 3.05 μW from 0.5 and 1.0 V supplies in an area of 0.05 mm2 with 63.8 dB signal-to-noise-and-distortion ratio and 3.02 noise efficiency factor
Raheem et al., 2018Bio-potentials2 channelsProgrammable gain amplifier (PGA) and 10-bitΣΔ (SDM-ADC)High impedance, power consumption of 11 mW, programmable gains from 52.6 dB to 72 dB and input referred noise of 3.5 µV in the amplifier bandwidth
Mazzetta et al., 2018EMGDifferential 1 channel32 bit ARM® Cortex®-M4, microSD, Bluetooth 4.0, 592 mWh battery, micro-USB connector, 30 × 30 × 15 mm dimensions, weight of 10 gPower consumption, compactness and energy autonomy, wireless and comfortably wearable
Biagetti et al., 2018sEMGN.A.Ultralight wireless sensing nodes, base station for data transmission through a 2.4 GHz radio link, communication protocol designed on top of the IEEE 802.15.4 physical layerLow-cost wearable wireless system, user interface software for viewing, recording and analyzing data
Kast et al., 2017Bio-potentialsBipolar 64 channelsUp to eight front-end acquisition modules with synchronization module, a separated universal serial bus data-link to the computer and an ADS1299Raw data are analyzed and stored on a personal computer or a single-board computer
Sarker et al., 2017ECG/EMG8 channels24 bit resolution/channel and 500 samples/s, IoT-based systemCompact and wearable portable bio-signal acquisition device, real-time data wireless transmission, low energy consumption
Li et al., 2017ECG/EMGN.A.150 mAh rechargeable Li-ion battery, packaged into a 39 × 32 × 17 mm 3D printed small box, total weight of 24.0 g, power management circuit, dual power supply for operational amplifiersWearable wireless non-contact system, ultra-high input impedance, feasibility of long-term biopotential monitoring
Senepati et al., 2017ECG/EMGN.A.Band pass and band stop FIR filters, Successive Approximation Register (SAR) DAC, Spartan-3E FPGA and 0.18 μm CMOS TSMC technologyArea of 33,005 μm2 area, power consumption of 0.382 mW, suppressing of baselines wander and power line interference noise (50/60 Hz)
Bhamra et al., 2017ECG/EMGN.A.ASIC technology in a 0.18 μm CMOS process, high-pass and low-pass cutoff frequencies being 0.5–300 Hz and 150 Hz–10 kHz, antialiasing filter, successive approximation register (SAR) analog-to-digital converter (ADC), power managementWireless, programmable gain from 38 to 72 dB, AFE and ADC dissipation of 5.74 μW and 306 nW, measured input-referred noise of 2.98 μVrms, noise efficiency factor of 2.6, power efficiency factor of 9.46, area of the AFE of 0.0228 mm2
Kim et al., 2016Bio-potentials, PPG, BIAN.A.CMOS technology, low-power and multimodal analog front-end (AFE)Wearable health monitoring, low dimension and power consumption
Mahmud et al., 2016ECGN.A.Fully integrated analog front-end (AFE), temperature sensor, accelerometer, Bluetooth Low Energy (BLE) moduleMultiparameter real time monitoring, small dimensions, Android application, alerts
Piccinini et al., 2016ECG/EMGN.A.ADS1294 Medical Analog Front End, CC3200 microcontroller, two Li-ion charged batteriesPortable solution, size physical reduction, robustness in wireless transmission, reliability in data acquisition and processing
Lee et al., 2016ECG/EMGN.A.Mixed-signal processor system-on-chip (SoC), Bluetooth Low Energy (BLE) chip, 200 mAh batteryWireless transmission, power efficiency, 12 h of continuous recording
Augustyniak et al., 2016Bio-potentialsSingle-ended 5 channelsProgrammable AFE ADAS1000, 24-bit resolution analog-to-digital converter with programmable data rate up to 128 kHzWired and wireless body sensor networks, configurable gain for channel

3. Commercial Wearable Devices

Wearable portable systems aim to daily acquire and processes different health data, providing early detection of pathological signs and improving the treatment and the continuous monitoring of disease. Many commercial EMG and ECG sensors are available, and they are designed and created to satisfy different specifications. In this section, the review proposes a description of the common commercial biosignal acquisition systems for physiological monitoring. These systems have been chosen to be the most used devices in health practice presenting similar characteristics to be compared.

Biometrics Ltd offers different data acquisition systems to collect analog and digital data from various sensors and are available in wireless, portable and laboratory configurations. Wireless systems furnish total freedom of movement without being constrained by wires [104]. They are available in 2-, 4-, 8- and 16-channel configurations to acquire EMG signals by using surface, small and lightweight sensors, allowing muscle activity readings to be smooth and robust with a range of up to 30 m from its receiver. The main features of these types of sensors are (i) a bandwidth from 10 Hz to 250 Hz through to 10 Hz to 5000 Hz and (ii) a sensitivity for the peak to peak measurements ranging from +/− 60 mV to +/− 6000 mV [105]. Portable systems are comprehensive packages of sensors and instrumentation for static and dynamic measurements in a clinical setting, a research centre, or at any remote location such as an office, workplace or home. Biometrics offers three different versions of EMG sensors: (i) surface EMG sensors, (ii) wireless surface EMG sensors and (iii) surface EMG amplifier.

Biosignalsplux represents an advanced wireless toolkit to collect and analyze reliable and high-definition biosignal data [112]. It offers a set of cabled and wearable sensors. The biosignalsplux electromyography (EMG) sensor is a high-performance bipolar sensor with low noise for seamless muscle data acquisition. This sensor is designed to monitor muscular activity, and the bipolar configuration is ideal for uncompromised low-noise data acquisition. The raw data output provides medical-grade data enabling it to be used for advanced and highly accurate biomedical biomechanics and sports research. Its main features are (i) bipolar differential measurement, (ii) pre-conditioned analog output, (iii) high signal-to-noise ratio and (iv) medical-grade raw data output. It is also ready-to-use, and it is miniaturized. The wireless single-channel EMG device for real-time muscle sensing is muscleBAND. It is an integrated single-channel EMG sensor with a triaxial accelerometer and magnetometer for real-time acquisition of muscle activity and motion data with an integrated dual Bluetooth module. This sensor allows data acquisitions with up to 16-bit resolution at up to 1000 Hz sampling rate, with the internal battery providing enough power for continuous data streaming.

Delsys proposes complete wireless EMG-based solutions for monitoring human movement in research, clinical and educational settings [113]. These solutions are composed of (i) research, mobile and lite systems, (ii) EMG sensors, (iii) mobile software and (iv) software for devices integration. The most used EMG sensor is Trigno Avanti Sensor, which can capture muscle activity and movement data accurately. It is designed to work with all Trigno systems, and it is characterized by (i) patented technology, (ii) improved RF performance, (iii) cable-free design, (iv) selectable EMG bandwidth settings and (v) on-board signal processing. It also allows differential EMG input acquisition in a very small dimension and weight. Trigno Research+ is a high-performing device designed to make EMG signal detection reliable and easy, offering a full set of physiological and biomechanical monitoring tools to simplify complex research and provide the highest quality data. Proprietary RF protocol guarantees synchronization between all sensors and allows data transmission from Trigno wireless sensors to a Trigno base station. Table 32 reports the main characteristics of the selected wearable monitoring systems.

Table 2. Main characteristics of the commercial wearable monitoring systems.
FeaturesBiometricShimmerBiosemiBTS BioengineeringBiosignal PluxBITalinoDelsys
Type of sensorWireless EMG SensorShimmer3 EMG UnitActiveTwoFreeEMG 1000 H2OElectro-myography SensorElectro-myography SensorTrigno Avanti Sensor
Size (mm × mm × mm)42 × 24 × 1465 × 32 × 12120 × 150 × 190Probes: 41.5 × 24.8 × 1428 × 70 × 1212 × 2727 × 37 × 13
Weight17 g31 g1.1 kg13 g—battery included25 gN.A.14 g
# channels128 up to 2561111 differential input
Input impedance>100 MohmsN.A.>100 M @ 50 Hz >100 GOhm10/7.5 GOhm/pF 
Input range+/−6 mVApprox. 800 mV @ gain = 6+262 mV to −262 mVN.A.Up to  10 mV±1.64 mV @ VCC = 3.3 V11 mV/22 mV rti
Gain+/−60 mV to +/−6000 mV1,2,3,4,6,8,12 (software configurable)N.A.N.A.1000100911 mV/22 mV rti
CMRR>96 dB (typically 110 dB) @ 60 HzN.A.>90 dB @ 50 HzN.A.100 dB86 dB<−80 dB
ConsumptionN.A.N.A.4 Watt @ 280 channelsN.A. 1 mA 0.17 mAN.A.
Bandwith0–250, 470, 950, 5000 Hz8.4 kHzUp to DC—3200 Hz @ –3 dBN.A.25–500 Hz25–482 Hz10–850 Hz 20–450 Hz
Data transmissionWirelessBluetooth Radio – RN-42Fiber opticWireless IEEE 802.15.4Bluetooth Low EnergyN.A.2.400-2.483 GHz ISM Band, Proprietary RF Protocol - BLE V4.2
ResolutionN.A.24 bit24 bit16 bit12 bitN.A.16 bit
Sample rateN.A.125, 250, 500, 1000, 2000, 4000, 8000 SPS2048 Hz–4096 Hz–8192 Hz–16,384 HzN.A.N.A.N.A.4370 sa/sec
Battery type and lifeRechargeable Li-ion Polymer, Up to 8 h450 mAh rechargeable Li-ion batteryBattery power with >10 h @ 144 channels, >72 h @ 16 channelsBattery Li-Po, Up to 6 hN.A.Battery Li-Po 700 mAhRechargeable Li-Po Battery Up to 8 h