Hydrogen Peroxide and Dental Environment: Comparison
Please note this is a comparison between Version 3 by Lindsay Dong and Version 2 by Lindsay Dong.

Hydrogen peroxide is an effective biocide in its gaseous (vaporized and aerosolized) form against viruses, spores, fungi, and bacteria [1,5]. The vaporized solution of hydrogen peroxide, which is based on water, is activated by plasma and acts as an oxidizing and disinfecting agent when it settles and contacts the surfaces of all objects in the room.

Hydrogen peroxide is an effective biocide in its gaseous (vaporized and aerosolized) form against viruses, spores, fungi, and bacteria. The vaporized solution of hydrogen peroxide, which is based on water, is activated by plasma and acts as an oxidizing and disinfecting agent when it settles and contacts the surfaces of all objects in the room.

  • hydrogen peroxide
  • vaporized hydrogen peroxide
  • bio-decontamination
  • enumeration

1. Introduction

Infection control has always been a core objective in dentistry; however, it has risen to greater importance given the SARS-CoV-2 pandemic. Aerosolized viruses and bacteria, such as Tuberculosis, Candida auris, and Staphylococcus aureus, once inhaled by either patient or healthcare worker, can result in far-reaching health consequences [1][2][3].

Infected droplets can be spread by dental instruments from the mouth of the dental patient, such as high-speed rotating handpieces and ultrasonic devices. The contaminated aerosol settle on exposed surfaces, resulting in environmental contamination. Despite surface disinfection protocols, many of the inanimate objects are not routinely disinfected and together with the hands of staff become vectors for transmission of healthcare-associated infections [3][4]. Studies determined that established disinfection methods showed their inability to eliminate environmental contamination of certain pathogens that are associated with direct transmission [5]. In addition, an increase in the bacterial load after the use of a neutral detergent was reported. A neutral detergent is composed of surfactants, fillers, and chelating agents. They are typically not used in pathogen eradication [6]. Enhanced cleaning that deviates from traditional surface cleaning (traditional surface cleaning or terminal cleaning aims to reduce the number of pathogens on surfaces to reduce transmission) not only reduced the bacterial load in the environment but also reduced the number of organisms on the hands of staff. Disinfection procedures that involve physical contact with the surfaces (spray, wipe, and spray techniques) are widely used but are usually labor-intensive and not always effective, as it is impossible to reach all hidden surfaces. For this reason, it is imperative to investigate the efficacy and then adopt other infection control approaches to decontaminate the dental environment between patients and minimize the risk of transmission of diseases [5][7].

VHP generators are no-touch decontamination and therefore circumvent problems associated with operators during manual disinfection such as incorrect application and use of cleaning agents.

2. Vaporized Hydrogen Peroxide as a Non-Contact Decontamination System for Pathogens Associated with the Dental Environment

The characteristics of VHP decontamination are presented in Table 1 and Table 2. Bioquell was the manufacturer of 13 of the 19 machines assessed in the selected studies [8][9][10][11][12][13][14][15][16][17][18][19][20]; however, irrespective of the VHP unit used, all the studies reported favorable outcomes (towards the VHP generators rather than the aerosolized hydrogen peroxide (aHP) generators) for log reduction of the assessed pathogens. Methicillin-resistant Staphylococcus aureus (MRSA) was the most used bacterial pathogen in five studies and with the viruses Feline calicivirus, Human norovirus, and Murine norovirus featured in three studies.

Most of the VHP generators are equipped with monitoring systems, but part per million (ppm) monitoring is essential to ensure the desired concentration of hydrogen peroxide is reached for the desired dwell time. Additionally, the use of standardized validated Geobacillus stearothermophilus biological indicators [9][10][11][12][15][16][17][18][21], are important to set the benchmark for the efficacy of the VHP with the manufacturer’s instructions. The surface that received the pathogen was predominantly stainless steel in the form of discs, coupons, or tape with the exceptions of cell culture well plates [8][11][12] and cryogenic tube caps [14]. Two authors used stainless steel as well as some additional materials [22][20]. The log kill was sufficient for all the authors to conclude that VHP generation was effective for the assessed pathogens. The studies that assessed aHP found a greater log kill with VHP generators [9][10].

Table 1. The characteristics of the included studies for VHP decontamination (n = 17).

Characteristics n or n (%)
Publication Year n = 17
2010 1 (0.05)
2011 3 (0.17)
2012 3 (0.17)
2014 1 (0.05)
2015 1 (0.05)
2016 2 (0.11)
2017 3 (0.17)
2019 2 (0.11)
2020 1 (0.05)
Location n = 17
United Kingdom 8 (0.47)
Sweden 3 (0.17)
USA 2 (0.11)
Brazil 1 (0.05)
France 1 (0.05)
Germany 1 (0.05)
The Netherlands 1 (0.05)
Hydrogen Peroxide Vapourizing machine (n = 19) or n% of total machines
Aeroclave 1 (0.05)
Bioquell 1 (0.05)
Bioquell BQ-50 1 (0.05)
Bioquell Clarus C 2 (0.10)
Bioquell Clarus L 1 (0.05)
Bioquell Clarus R 3 (0.15)
Bioquell Clarus S 1 (0.05)
Bioquell Q10 4 (0.21)
Liquid Verne Veiling equipment 1 (0.05)
Sterinis aHP 1 (0.05)
Steris La Calhene VHP 1 (0.05)
Steris VHP 1 (0.05)
Sterinis system SR2 1 (0.05)
Assessed Pathogen: n or n% of total pathogens
Candida (n = 34)
Various Candida species 34 (100)
Bacteria (n = 27)
Acholeplasma laidlawii 1 (0.03)
Acinetobacter baumannii 1 (0.03)
Bacillus anthracis (Ames) spores 1 (0.03)
Brucella abortus 1 (0.03)
Burkholderia pseudomallei 1 (0.03)
Clostridium difficile 1 (0.03)
Escherichia coli 1 (0.03)
Geobacillus stearothermophilus biological indicators 9 (0.33)
MDR Acinetobacter baumannii 1 (0.03)
Methicillin-resistant Staphylococcus aureus (MRSA) 5 (0.18)
Mycoplasma pneumoniae 1 (0.03)
Mycoplasma gallisepticum 1 (0.03)
Mycobacterium tuberculosis 1 (0.03)
Vancomycin- resistant Enterococcus (VRE) 1 (0.03)
Yersinia pestis 1 (0.03)
Virus n = 21
Adenovirus 2 (0.09)
Avian influenza virus (AIV) 1 (0.04)
Escherichia virus MS2 1 (0.04)
Feline Calicivirus 3 (0.14)
Foot and mouth disease (FMDV) 1 (0.04)
Human adenovirus type 1 1 (0.04)
Human norovirus 3 (0.14)
Influenza A virus (H1N1) 1 (0.04)
Murine norovirus (MNV) 3 (0.14)
Pseudomonas virus phi6 1 (0.04)
Poliovirus 1 (0.04)
Rotavirus 1 (0.04)
Swine influenza virus (SwIV) 1 (0.04)
Transmissible Gastroenteritis coronavirus of pigs (TGEV) 1 (0.04)
Characteristics:  
Method of inoculation n = 29, n% of total surfaces
Sabouraud’s dextrose agar and fabric 2 (0.06)
Stainless steel 10 mm-diameter discs/coupon 5 (0.17)
Stainless steel 3 mm-diameter discs/coupon 1 (0.03)
Stainless steel 2.2 cm × 2.5 cm disc 1 (0.03)
Tyvek-pouched stainless steel disc/coupon 5 (0.17)
Plastic plates 2 (0.06)
Steel embossing tape 2.5 cm × 5 cm 1 (0.03)
Roller bottle 1 (0.03)
Unspecified stainless steel/coupon 4 (0.14)
Gauze 1 (0.03)
Glass 1 (0.03)
Painted joint tape 1 (0.03)
Wood 2 (0.06)
Ceramic tile 1 (0.03)
N95 Filter medium 1 (0.03)
Efficacy: Log kill  
>8 log 1
>6 log 4
>4 log 4
>3 log 2
<2 log 2
1.3–3.5 log reduction 1
Log reduction not specified 4

Table 2. Summary of items regarded in the risk assessment of the chosen article.

Author Country of Study Aim/Objective Pathogen Used Methodology:-Hydrogen Peroxide Concentration Blinding and Controls Sample Hanling and Contamination Prevention Failed Experiments and Data/Results Not Presented Pathogens Placed on Material Outcome (Level of Bio-Decontamination)
A Abdolrasouli et al. 2017 United Kingdom In vitro evaluation of the efficacy of VHP on standard and outbreak C. auris. 34 different yeast isolates: 4 strains (Candida albicans,

Candida tropicalis, Candida krusei, Candida parapsilosis) 28 outbreak isolates of C. auris
Bioquell machine, No H2O2 liquid concentration, performed following manufacture instructions. 8 g of H2O2/m3. No blinding. One C. aurus control plate with no exposure to VHP. Six yeast-free control wells. No BI used. Done in triplicate. Wells of pathogen grown in a 96-well plate and desiccated, sealed, and kept at 4 °C until exposure to VHP. Viability was then assessed on SDA with C. auris control plate. Did not state how long after fogging well plates were closed to prevent contamination. One Indian C. auris and a specific Indian strain not named. Non-exposed C. auris. Candida species and VHP exposed C. aurus survive in a desiccated state. Data not shown. Well plates. Data provided evidence

that C. auris (and other Candida species) are effectively killed with a 96.6–100% by H2O2 vaporization.
E Berrie et al. 2011 United Kingdom In vitro efficacy of inactivation of recombinant adenovirus by VHP. Dried recombinant adenovirus

(Ad5GFP)
Bioquell Clarus S machine, 60 mL of 30% H2O2 liquid concentration, performed dwell time 45 min. Whole VHP cycle 3 h Exposed and non-exposed samples to VHP. BI (Biological Indicator) indicators used. The experiment is one disc per dilution and repeated in triplicate Immediately after the experiment, the samples were transferred to a sterile microbiological safety cabinet. One to two logs of pathogen lost due to drying or recovery method compared to wet reference samples compared to experiment two at the titer. Viability reduction data explained in the article. Stainless steel 10-mm-diameter discs. Data provided evidence

that Adenovirus are effectively killed with a 7.6 to 9.4 log kill by H2O2 vaporization.
M Eterpi et al. 2010 France & United Kingdom In vitro evaluation of the efficacy

of VHP and cold VHP sterilization against Mycoplasma.
Mycoplasma gallisepticum, M. pneumoniae, and A. laidlawii VHP100 Steris machine. 30% H2O2 liquid concentration. Three cycles 1200 ppm/15 min; 400–500 ppm/60 min; 180–200 ppm/4 h. No blinding. VHP unexposed samples kept under a laminar flow hood in sealed Petri dishes for the same time cycle and managed the same as exposed coupons. No BI used. Six treated samples with each method and repeated for times. Samples were transferred to an SP4-glucose broth immediately after VHP exposure. Less than one log of pathogen lost due to drying or recovery method as described by Nagatomo et al. 2001 with loss due to recovery ≤0.5 log. Neutralization an additional ≤0.5 log. Viability reduction data explained in the article. Stainless steel coupons of 1 cm × 3 cm. Data provided evidence

that Mycoplasma is effective with a >4 log kill by H2O2 vaporization.
T. Y. Fu et al. 2011 United Kingdom Compare the efficacy, efficiency of VHP and aHP. Methicillin-resistant Staphylococcus aureus (MRSA), Clostridium difficile and Acinetobacter baumannii. Bioquell Clarus R machine. 30% H2O2 liquid. SR2 Sterinis machine with a 5% H2O2 liquid and silver ion (50 ppm) and orthophosphoric acid (<50 ppm), dose 6 mL/m3 recommended by the manufacturer. No blinding. Both exposed and non-exposed to VHP. BI used. Four cycles per machine with each cycle consisting of three unexposed VHP/aHP and three dry VHP/aHP discs for water, 3% BSA (Bovine Serum Albumin) and 10% BSA. The control was cycled separately over four cycles. Did not state how long after fogging discs were transferred to prevent contamination, nor the overnight drying to prevent contamination. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel discs with a diameter of 10 mm. The VHP system achieved a greater level of biological inactivation between 4–6 log for most locations than the aHP system 1–5 log depending on the pathogen.
Goyal et al. 2014 United Kingdom Evaluate the in vitro efficacy of three volumes of VHP on selected viruses with surface contamination. FCV as a

surrogate of human norovirus, TGEV as a surrogate

for the SARS virus,

human adenovirus type 1, AIV (A/chicken/Maryland/

2007[H9N9]) and SwIV (A/swine/Minnesota/2010 [H3N2]).
Bioquell Clarus L machine. 35% H2O2 liquid. Hydrogen peroxide at 2 mL/min for 1, 2, 5 min followed by 1.5 mL/min or 15 min equating to the following different volumes: 25, 27, and 33 mL with the treatment time between 2–3 h for the completed cycle. No blinding. Non-VHP exposed inoculated discs at room temperature. Four BI were exposed to the VHP in corners of the environmental chamber. Positive BI control was not exposed to VHP. 8% Fetal Bovine Serum (FBS) served as soiling present in the culture medium. Each experiment had inoculated discs exposed to each vaporized volume of VHP and one disc not exposed to VHP. Discs are left to dry in a biosafety cabinet to prevent contamination. After VHP the discs including the non-exposed control discs were transferred immediately to the environmental chamber for titration. Data was determined concerning the control disc. This allows direct comparison to the test and control discs having the same log loss, making the comparison more accurate. But also leads to not knowing what the log loss of viral load is. Loss of virus log particles during the methodology of drying and recovery. Not calculated 10 mm stainless steel discs. VHP was virucidal for viruses assessed dried on surfaces, suggesting that VHP can be considered for the disinfection of virus-contaminated surfaces based on the 8% FBS surface contamination.
Holmdahl et al. 2011 Sweden Comparison of VHP and aHP to BI in various locations. BI with G. stearothermophilus Steris VHP machine. 5% H2O2 liquid. 6 mL/m3 with 100–150 ppm. Bioquell Q10 machine. 35% H2O2 liquid. 900 mL per test and results in 6.6 g/m3. 338 ppm peak, 3 h. No blinding. BI was used as control. BI in Tyvek pouches Direct comparison of two machines on BI. All results presented. BI stainless steel disc placed in various locations in the room. All results presented for the same areas assessed for the two machines. VHP showed a 100% negative result while aHP presented with multiple positive results, the inconsistency with the aHP was 10% kill (100 ppm) followed by two cycles of 79% kill, with the ppm in cycles 2 and 3 being 130 and >150 ppm respectively.
Holmdahl et al. 2016 Sweden Evaluate the efficacy of VHP in six locations for two virus pathogens with surface contamination. FCV, feline permissive cell line (FCWF). MNV and permissive murine cell line (RAW 264.7) Bioquell Q10 machine. As per the manufacturer. No H2O2 liquid concentration. Gassing time 40–50 min, dwell time 15 min. VHP ppm range 474–505 ppm with a total cycle time of 3 h. Virus prepared in triplicate in well plates. Two inoculated plates and BI not VHP exposed two areas of the control room. BI exposed at all the positions with VHP next to virus inoculated plates. Each VHP exposure experiment was repeated in triplicate. Well plates left to dry at room temperature under a hood for 2 h and stored. Loss of virus log particles during the methodology of drying and recovery were calculated. Well plates VHP was virucidal for viruses assessed dried on surfaces, suggesting that VHP can be considered for the disinfection of virus-contaminated surfaces based on the 10% FBS surface contamination.
Holmdahl et al. 2019 Sweden Assess norovirus viability of cytopathic

the effect after VHP.
Two human norovirus field strains, genogroup I and II. Murine norovirus. Bioquell Q10 machine. No H2O2 liquid concentration. 860 ppm VHP for 33 min gassing and 55 min dwell. This resulted in 205 g of H2O2 used. No blinding. BI and mock samples with no VHP exposure. Virus samples dried in 35 mm diameter wells of six-well plates mock and VHP treated samples. Data was determined concerning the lowest detection limit of 10−0.5. This allows direct comparison to the test and control discs having the same log loss, making the comparison more accurate. But also leads to not knowing what the log loss of viral load is. Well pates BI deactivated and norovirus log 5 kill.
Lemmen et al. 2015 Germany Efficacy of VHP on five pathogens dried onto various hard surfaces. MDR MRSA and MDR VRE, MDR A baumannii. BI as proxy for D. difficile Bioquell Q10 machine.

30% H2O2 liquid. Three cycles were performed. The dose of 11.2 g/m3 achieved after 50–52 min until hydrogen peroxide was 500–600 ppm. 20 min dwell time.
No blinding. BI used. Four of each material inoculated with the pathogen and distributed in four locations exposed to VHP and the same number not exposed to VHP as controls. BI placed in 4 corners of the room and 3 challenge locations. Kept on a sterile basis until experiment and after VHP exposure transferred to a sterile glass tube with 1 mL distilled water. Lost pathogens are known and presented in the article and mean log reduction is calculated. Stainless steel discs, gauze VHP inactivated all spore BI (>6 log10 reduction), and no MRSA, VRE, or MDR A baumannii were recovered from the stainless steel and cotton carriers (>4–5 log10 reduction, depending on the starting inoculum). VHP was equally effective at all carrier locations. No difference in efficacy
Montazeri et al. 2017 USA Inactivation of human norovirus after VHP exposure. FCV. Outbreak human NoV GI.6 and GII.4. AeroClave System 3110. 7.5% H2O2 liquid. No air handling unit during vapor process, at end of cycle turned on for 20 min. 7.1–15.9 mL/m3 was achieved, with 5 min dwell time following the manufacturer’s recommendation. Then the air handling unit was switched on for 20 min. No blinding. No BI used. 7 locations in BSL-3 laboratory assessed with VHP. Inoculated coupons not exposed to VHP were outside the laboratory for the duration of the experiment. Air-dried in a biosafety hood. And used immediately for the experiment. After the experiment, the samples were transferred to PBS tubes. Data was determined to the control disc. This allows direct comparison to the test and control discs having the same log loss, making the comparison more accurate. But also leads to not knowing what the log loss of viral load is. Stainless steel embossing tape No trend was observed for human NoV GI.6 reduction as a function of H2O2-based disinfectant formulation concentration. However, increasing the concentration from 7.1 to 12.4 mL/m3 enhanced viral genomic copy number reduction for GII.4
Murdoch et al. 2016 United Kingdom Assess the application of three different liquid concentrations for VHP. MRSA and Geobacillus stearothermophilus Bioquell BQ50 machine. 5, 10, and 35% H2O2. 640 g hydrogen peroxide over 40 min and 200 min dwell time. No blinding labeled containers. BI used. Positive and negative controls. Every 10 min throughout the experiment a BI was exposed for 10 min. All specimens were placed in labeled 30 mL containers. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel discs 35% hydrogen peroxide is ideal.
Otter et al. 2012 United Kingdom Efficacy of VHP against methicillin-resistant Staphylococcus aureus on various surfaces. MRSA Bioquell Clarus R machine. No H2O2 liquid concentration. VHP mean concentration

134 ppm.
No blinding. No BI used. Control discs were not VHP exposed. The experiment ran in triplicate per period for each contaminant material. Air-dried in the test room air, then VHP exposure and immediately enumerated. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel discs Relative susceptibility to VHP was 10% BSA < TSB < 3% BSA = water. At a ppm achieved and >75 min exposure, no MRSA was recovered on the discs.
Petit et al. 2017 Brazil Efficacy of VHP against foot-and-mouth disease. Three serotypes of Foot-and-mouth disease virus (FMDV) Bioquell Clarus R machine. 35% H2O2 liquid. 115 min. VHP injection time 75 min, 40 min dwell time. No blinding. No validated BI manufactured by VHP producers. Positive controls of three serotypes. Three replicate cycles of 15 BI produced from FMDV for VHP exposure. Five samples for each viral serotype were produced. One plosive control per serotype for the duration of the experiment was stored in a refrigerator. Dried in class 2 biological safety cabinet. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Inside the cap of the polypropylene cryogenic tube. Three FMDV serotypes showed full inactivation.
Pottage et al. 2012 United Kingdom Comparison of log kill of BI vs MRSA after VHP exposure. G. stearothermophilus and MRSA A Steris VHP-1000ARD machine. 35% H2O2 liquid. 750 ppm maintained in chamber. No blinding. Random removal of VHP exposed samples. 18 MRSA and 18 BI indicators placed in sterile Petri dishes and VHP exposed for pre-determined periods. Three unexposed stainless steel discs of each pathogen were used. Inoculated stainless steel discs air-dried for 1 h. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. BI on stainless steel discs sealed in Tyvek packages. BI greater log kill than MRSA for the same periods of exposure.
Pottage et al. 2019 United Kingdom Efficacy of VHP on dried bacteria. Bacillus anthracis (Ames) spores, Brucella abortus,

Burkholderia pseudomallei, Escherichia coli, Mycobacterium

tuberculosis and Yersinia pestis.
Bioquell Clarus C machine. 35% H2O2 liquid. 90 min cycle. No blinding. Three controls tied in double plastic bags to determine the loss of log pathogen. 3 control samples were used as the start pathogen load. 12 produced BI for each VHP run to allow triplicate exposure. Three control BI from a VHP manufacturer used per VHP cycle. Dried in a biological cabinet for 1 h. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel coupons in Petri dishes This study demonstrates that VHP can inactivate a range

of HG3 agents at high concentrations with associated organic matter, but M.

tuberculosis showed increased resistance to the process.
Tuladhar et al. 2012 The Netherlands Virucidal efficacy of VHP against respiratory and enteric viruses on various materials. Poliovirus, human norovirus genogroup II.4 (GII.4), murine norovirus 1, rotavirus, adenovirus, and influenza A (H1N1) virus. Boneco 7131 machine. 12% H2O2 liquid. 120–134 ppm at a flow rate of 2.3 L/h. No validated BI manufactured by VHP producers. Triplicate samples per virus were performed twice. Control samples were not VHP exposed. Dried in a biohazard cabinet. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel, framing panel, and gauze carriers. VHP effective against pathogens assessed.
Wood et al. 2020 USA Assess the decontamination efficacy of VHP on phages. Bacteriophage viruses, MS2 and

Phi6
Humidifier with 3 or 8% H2O2 liquid generated to 25 ppm. Bioquell Clarus C machine. 35% H2O2 liquid. 25 ppm and 400 ppm generated. No validated BI manufactured by VHP producers. Inoculated samples, not VHP exposed, and inoculated samples VHP exposed. Two blank samples. Completed in triplicate. Samples made and dried in a biosafety cabinet. After the experiment coupons were sealed and transferred to the biosafety cabinet. No pathogens lost or contaminated samples were described or considered in the methodology. All data presented. Stainless steel, glass, tile, N95 mask material, painted joint tape, wood. Extrapolating from these results for both an enveloped and non-enveloped virus, we would expect LCHP would be a viable decontamination option for EBOV for relatively clean surfaces

3. Summary

The overarching conclusion is that H2O2 delivered as VHP was an effective method to achieve large levels of log kill on the assessed pathogens. All the articles have applications to dentistry bio-decontamination. They showed the efficacy of VHP in spaces and surfaces similar to a dental clinic. Further investigation of VHP in dental clinics is required with certain variables that must be known and standardized to assure the validity and reproducibility regarding the H2O2 concentration, dwell time, and a constant ppm or defined ppm range during the dwell time. The enumerated pathogens at every step of the methodology, from inoculation on the test surface to the enumeration of the exposed and unexposed samples, should be completed. This safeguard will ensure the correct determination of the log loss of pathogens. From the results of the reviewed articles, a statistically calculated sample size performed in triplicate should be standardized.

References

  1. Ali, S.; Muzslay, M.; Bruce, M.; Jeanes, A.; Moore, G.; Wilson, A.P.R. Efficacy of two hydrogen peroxide vapour aerial decontamination systems for enhanced disinfection of meticillin-resistant Staphylococcus aureus, Klebsiella pneumoniae and Clostridium difficile in single isolation rooms. J. Hosp. Infect. 2016, 93, 70–77.
  2. Schelenz, S.; Hagen, F.; Rhodes, J.L.; Abdolrasouli, A.; Chowdhary, A.; Hall, A.; Ryan, L.; Shackelton, J.; Trimlett, R.; Meis, J.F.; et al. First hospital outbreak of the globally emerging Candida auris in a European Hospital. Antimicrob. Resist. Infect. Control 2016, 5, 35.
  3. Scarano, A.; Inchingolo, F.; Lorusso, F. Environmental Disinfection of a Dental Clinic during the Covid-19 Pandemic: A Narrative Insight. BioMed Res. Int. 2020.
  4. Otter, J.A.; Mepham, S.; Athan, B.; Mack, D.; Smith, R.; Jacobs, M.; Hobkins, S. Terminal decontamination of the Royal Free London’s high-level isolation unit after a case of Ebola virus disease using hydrogen peroxide vapour. Am. J. Infect. Control 2015, 44, 233–235.
  5. Falag, M.E.; Thomaidis, P.C.; Kotsantis, I.K.; Sgouros, K.; Samonis, G.; Karageorgopoulos, D.E. Airborne hydrogen peroxide for disinfection of the hospital environment and infection control: A systematic review. J. Hosp. Infect. 2011, 78, 171–177.
  6. Chan, H.T.; White, P.; Sheorey, H.; Cocks, J.; Waters, M.-J. Evaluation of the biological efficacy of hydrogen peroxide vapour decontamination in wards of an Australian hospital. J. Hosp. 2011, 79, 125–128.
  7. Tysiąc-Miśta, M.; Dubiel, A.; Brzoza, K.; Burek, M.; Pałkiewicz, K. Air disinfection procedures in the dental office during the covid-19 pandemic. Med. Pracy 2021, 72.
  8. Abdolrasouli, A.; Armstrong-James, D.; Ryan, L.; Schelenz, S. In vitro efficacy of disinfectants utilised for skin decolonisation and environmental decontamination during a hospital outbreak with Candida auris. Mycoses 2017, 60, 758–763.
  9. Fu, T.Y.; Gent, P.; Kumar, V. Efficacy, efficiency and safety aspects of hydrogen peroxide vapour and aerosolized hydrogen peroxide room disinfection systems. J. Hosp. Infect. 2012, 80, 199–205.
  10. Holmdahl, T.; Lanbeck, P.; Wullt, M.; Walder, M.H. A head-to-head comparison of hydrogen peroxide vapor and aerosol room decontamination systems. Infect. Control Hosp. Epidemiol. 2011, 32, 831–836.
  11. Holmdahl, T.; Walder, M.; Uzcátegui, N.; Odenholt, I.; Lanbeck, P.; Medstrand, P.; Widell, A. Hydrogen peroxide vapor decontamination in a patient room using feline calicivirus and murine norovirus as surrogate markers for human norovirus. Infect. Control Hosp. Epidemiol. 2016, 37, 561–566.
  12. Holmdahl, T.; Odenholt, I.; Riesbeck, K.; Medstrand, P.; Widell, A. Hydrogen peroxide vapour treatment inactivates norovirus but has limited effect on post-treatment viral RNA levels. Infect. Dis. 2019, 51, 197–205.
  13. Otter, J.A.; Yezlia, S.; French, G.L. Impact of the suspending medium on susceptibility of meticillin-resistant Staphylococcus aureus to hydrogen peroxide vapour decontamination. J. Hosp. Infect. 2012, 82, 213–215.
  14. Petit, B.M.; Almeida, F.C.; Uchiyama, T.R.; Lopes, F.O.C.; Tino, K.H.; Chewins, J. Evaluating the efficacy of hydrogen peroxide vapour against foot-and-mouth disease virus within a BSL4 biosafety facility. Lett. Appl. Microbiol. 2017, 65, 281–284.
  15. Goyala, S.M.; Chandera, Y.; Yezlib, S.; Otter, J.A. Evaluating the virucidal efficacy of hydrogen peroxide vapour. J. Hosp. Infect. 2014, 86, 255–259.
  16. Lemmen, S.; Scheithauer, S.; Häfner, H.; Yezli, S.; Mohr, M.; Otter, J.A. Evaluation of hydrogen peroxide vapor for the inactivation of nosocomial pathogens on porous and nonporous surfaces. Am. J. Infect. Control 2015, 43, 82–85.
  17. Murdoch, L.E.; Bailey, L.; Banham, E.; Watson, F.; Adams, N.M.T.; Chewins, J. Evaluating different concentrations of hydrogen peroxidein an automated room disinfection system. Lett. Appl. Microbiol. 2016, 63, 178–182.
  18. Berrie, E.; Andrews, L.; Yezli, S.; Otter, J.A. Hydrogen peroxide vapour (HPV) inactivation of adenovirus. Lett. Appl. Microbiol. 2011, 52, 555–558.
  19. Pottage, T.; Lewis, S.; Lansley, A.; Fraser, S.; Hendon-Dunn, C.; Bacon, J.; Ngabo, D.; Parks, S.R.; Bennett, A.M. Hazard Group 3 agent decontamination using hydrogen peroxide vapour in a class III microbiological safety cabinet. J. Appl. Microbiol. 2019, 128, 116–123.
  20. Wood, J.P.; Richter, W.; Sunderman, M.; Worth-Calfee, M.; Serre, S.; Mickelsen, L. Evaluating the Environmental Persistence and Inactivation of MS2 Bacteriophage and the Presumed Ebola Virus Surrogate Phi6 Using Low Concentration Hydrogen Peroxide Vapor. Environ. Sci. Technol. 2020, 54, 3581–3590.
  21. Pottage, T.; Macken, S.; Walker, J.T.; Bennett, A.M. Meticillin-resistant Staphylococcus aureus is more resistant to vaporized hydrogen peroxide than commercial Geobacillus stearothermophilus biological indicators. J. Hosp. Infect. 2012, 80, 41–45.
  22. Tuladhara, E.; Terpstrac, P.; Koopmansa, M.; Duizera, E. Virucidal efficacy of hydrogen peroxide vapour disinfection. J. Hosp. Infect. 2012, 80, 110–115.