Submitted Successfully!
To reward your contribution, here is a gift for you: A free trial for our video production service.
Thank you for your contribution! You can also upload a video entry or images related to this topic.
Version Summary Created by Modification Content Size Created at Operation
1 The soft tissue seal and the bone stimulation, preventing the initiation of peri-implantitis, depend on the biomechanics of implant-abutment connection. + 3876 word(s) 3876 2019-12-25 03:16:26 |
2 The soft tissue seal and the bone stimulation, preventing the initiation of peri-implantitis, depend on the biomechanics of implant-abutment connection. + 119 word(s) 3995 2019-12-29 01:12:35 | |
3 format change + 113 word(s) 3989 2019-12-30 04:14:47 | |
4 format change Meta information modification 3989 2020-07-31 05:25:10 | |
5 format change -21 word(s) 3968 2020-10-27 03:29:07 |

Video Upload Options

Do you have a full video?

Confirm

Are you sure to Delete?
Cite
If you have any further questions, please contact Encyclopedia Editorial Office.
Kim, J.; Lee, J.; Kim, J.C.; Lee, J.; Yeo, I.L. Tissues to Implant Abutments. Encyclopedia. Available online: https://encyclopedia.pub/entry/192 (accessed on 28 March 2024).
Kim J, Lee J, Kim JC, Lee J, Yeo IL. Tissues to Implant Abutments. Encyclopedia. Available at: https://encyclopedia.pub/entry/192. Accessed March 28, 2024.
Kim, Jung-Ju, Jae-Hyun Lee, Jeong Chan Kim, Jun-Beom Lee, In-Sung Luke Yeo. "Tissues to Implant Abutments" Encyclopedia, https://encyclopedia.pub/entry/192 (accessed March 28, 2024).
Kim, J., Lee, J., Kim, J.C., Lee, J., & Yeo, I.L. (2019, December 29). Tissues to Implant Abutments. In Encyclopedia. https://encyclopedia.pub/entry/192
Kim, Jung-Ju, et al. "Tissues to Implant Abutments." Encyclopedia. Web. 29 December, 2019.
Tissues to Implant Abutments
Edit

This entry is based on the fact that physical or biomechanical phenomena can cause biologic results. Implant-abutment connection structures determine the stability of soft tissue attachment to abutments. The bone responses to the strain that the stress is converted to, not the stress itself. Some implant-abutment materials and connection designs prevent the marginal bone loss by converting the stress to the strain stimulating bone apposition properly. These biomechanical interpretation of soft tissue seal and bone stimulation should be understood for the clinical long-term success of dental implants.

abutment implant-abutment connection soft tissue seal bone stimulation peri-implantitis marginal bone dental implant

1. Soft Tissue Responses to Different Implant System Materials and Structures

1.1. The Soft Tissue Seal Theory

Humans are exposed to a variety of external environments involving external forces, ultraviolet rays, and microorganisms. Human skin is the first line of defense that protects the human body against these external stresses. When human skin is pierced, the resulting hole must be closed by the immune system and healing mechanisms.

Teeth are one of the few organs in the human body that are located across the skin. The root of the tooth is surrounded by alveolar bone, while the part that penetrates through the soft tissue is in contact with epithelial tissue and/or the connective tissue of the mucosa[1]. Holes in the mucosa created by teeth are sealed by a special structure composed of epithelium and connective tissue[1]. An internal basal lamina and the hemi-desmosomes of epithelial tissue are attached to teeth, and a combination of dento-gingival fiber and cementum links teeth to the surrounding connective tissue. The holes connecting the inside and outside of the human body that contain teeth are secured by soft tissue attachments[1].

Like natural teeth, dental implants are an artificial organ located in a hole on the surface of the body, and the hole is sealed via a mechanism similar to the one that seals the holes around natural teeth using soft tissues. However, these sealing mechanisms are not identical to each other. Dental implants are in contact with the alveolar bone or soft tissue (epithelial and connective tissue) of the transition area (Figure 2)[2]. Fibers in the connective tissue attached to the abutment mainly run parallel to the surface and are circular in shape, whereas dento-gingival fibers, such as Sharpey’s fibers, are attached vertically to the cementum in natural teeth (Figure 1)[2].

Materials 13 00072 g002 550

Figure 2. Soft and hard tissues with collagen fibers surrounding a natural tooth (left) and a dental implant (right).

A scanning electron microscopy study showed that there are two distinct layers in the area within 200 μm of where the implant abutment comes into contact with the connective tissue. The inner 40 μm of this layer contains multiple fibroblasts that are attached directly to the surface of the abutment, while the outer 160 μm of this layer contains numerous collagen fibers[3]. Hence, the attachment of connective tissue to the abutment is maintained by fibroblasts attached to the titanium surface of the implant, and the elasticity of circular collagen fibers. For this reason, compared with the attachments of natural teeth, which involve direct attachment to connective tissue, attachments formed by connective tissue around an implant abutment are weaker.

Epithelial tissue is attached to the implant abutment via an internal basal lamina and hemi-desmosomes in a manner similar to the attachment of natural teeth[4][5][6][7]. However, epithelium attachment includes the internal basal lamina and hemi-desmosomes formed only in the lower part of the peri-implant epithelium around the implant abutment, whereas in natural teeth, these attachments are widely distributed throughout the junctional epithelium–tooth interface[8][9][10]. Thus, epithelial adhesions that form around the implant abutment are more vulnerable than those that form around natural teeth owing to their limited area of distribution.

In summary, both epithelial and connective tissues are more weakly attached to implant abutments than to natural teeth. Hence, holes on the surface of the human body that contain implants are more vulnerable to containment failure. This mechanism of blockade by soft tissue is called a ‘soft tissue seal’[8][11].

1.2. Attachment of Soft Tissue

The stability and immobility of soft tissue attachments in contact with the implant abutment are important factors that affect the long-term prognosis of the implant[8][11]. If the soft tissues around the implant are deformed owing to the movement of the lip, cheek, tongue, or jaw, the weak soft tissue seal surrounding the implant may be destroyed. This allows microbes to penetrate through the damaged mucosal seal, increasing the likelihood of disease around the implant[8][11]. Therefore, a stable soft tissue seal is essential to prevent microbial invasion and peri-implant disease[11][12][13].

The soft tissues around natural teeth are separated into the lining mucosa and masticatory mucosa, of which the masticatory mucosa is composed of the free and attached gingiva (Figure 3). The attached gingiva is in permanent and intimate contact with the surface of the enamel, the cementum, and the alveolar bone, thereby immobilizing the soft tissue[1]. It is, therefore, possible to retain the firmness and health of the mucosal seal by preventing it from detaching.

Materials 13 00072 g003 550

Figure 3. The structure of soft tissue around a natural tooth (left) and a dental implant (right).

Among the soft tissues surrounding an implant, the attached gingiva differs from those surrounding natural teeth[2]. The gingiva attached to the peri-implant can be divided into two parts: bone-attached and abutment (or implant)-attached gingiva (Figure 3). The bone-attached gingiva is identical to the corresponding tissue in natural gums, and rigidly immobilizes the soft tissue. However, as mentioned above, the abutment- (or implant)-attached gingiva more weakly adheres than the attachment gingiva that surrounds natural teeth[2]. Owing to the weak structure of the abutment- (or implant)-attached gingiva, the width of the bone-attached gingiva around the implant should be wide enough to prevent mucosal mobility, and thereby prevent the occurrence of peri-implant disease due to the collapse of the mucosal seal[2][11].

1.3. Disruption of the Soft Tissue Seal

The soft tissue seal is mainly destroyed via two mechanisms; instability of the peri-implant mucosa or the implant-abutment assembly. If there is no bone-attached gingiva around the implant, the soft tissue can become mobile, and the mucosal seal will inevitably rupture. When the mucosal seal is destroyed, bacteria can penetrate the internal environment through the transmucosal rupture site, potentially leading to peri-implant disease[14]. Therefore, because of the weakness of the soft tissue attachment, it is important to ensure there is sufficient bone-attached gingiva around the implant. Thus, a plan for implant surgery should be established to maintain the bone-attached gingiva following implant placement.

Instability in the connection between the implant and the abutment can also lead to the disruption of the soft tissue seal, which may present as mobility of the abutment or the implant. Mobility of the implant occurs only when osseointegration fails, but mobility of the abutment can occur as a result of various causes, including fractures of the abutment or implant, even following successful osseointegration. The most common cause of abutment mobility is the loosening of the abutment screw that connects it to the implant. Elongation and loosening of the screw caused by the lateral force of the masticatory load are more frequently observed in external connection-type implants than in internal connection-type implants[15].

In external-type implants, there is slight machining tolerance around the hex component, resulting in micromobility in the abutment. This external hex connection is, therefore, a mobile structure. Thus, most occlusal forces are concentrated on the abutment screw. This micromobility is likely to cause disruption of the soft tissue seal, bacterial infiltration, and peri-implant disease[16](Figure 4). Implant systems with an external hex connection were originally developed for the mandibular restoration of a completely edentulous patient wearing a maxillary complete denture. The occlusal force in such patients is weak, and this mobile connection is able to bear this weak masticatory force with few screw loosening events or breakdown of the soft tissue seal. However, such unstable implant–abutment connections can provoke severe peri-implant problems for a partially edentulous patient with antagonistic natural teeth and a strong occlusal force. To reduce the micromobility of the abutment, the external hex must be lengthened, and the machining tolerance between the male and female components must be minimized. Furthermore, it may be possible to fabricate an abutment screw using a stronger material, or by widening the platform size of the top of the implant to minimize the force concentrated on the abutment screw. In addition, when the abutment is fastened to the implant, applying higher torque may also be employed to increase the stability of the connection. However, none of these methods can completely eliminate the micromobility of the abutment in external connection-type implants.

In internal connection-type implants, the abutment–implant connection is firm owing to a process similar to cold-welding. Unlike external connections, in which the occlusal force is concentrated mainly on the abutment screw, the occlusal force is transmitted to the implant inner wall through the abutment–implant connection in internal connection-type implants. Therefore, less screw loosening occurs in this type of implant. In addition, the abutment–implant contact area is wider internal connection-type implants, and this prevents stress from concentrating at specific sites such as the abutment screw, and contributes to the stability of the soft tissue seal.

Materials 13 00072 g004 550

Figure 4. Schematic illustration of the ‘soft tissue seal’ theory. Micromobility of the abutment in external connection-type implants disrupts the surrounding soft tissue seal, which plays an important role in preventing external irritants from penetrating into the body. (a) External connection-type implant. (b) Parts experiencing the greatest stress (red lines) from eccentric forces (red arrows). (c) Abutment screw-loosening (red rotated arrow) by eccentric forces (red arrows).

1.4. Submerged and Nonsubmerged Implants

Depending on whether the top of the implant is at the alveolar bone or gingival level, implants can be divided into submerged-type and nonsubmerged-type. For submerged implants, the soft tissue seal is formed at the abutment area, rather than at the neck part for the nonsubmerged type. Histologically, there are no significant differences in the degree and pattern of soft tissue attachment to implants between submerged and nonsubmerged types[17]. When nonsubmerged implants are placed at the proper vertical position of the alveolar bone, the mobility of the abutment does not interfere with the soft tissue seal, or cause marginal bone resorption, because the interface between the abutment and the implant is positioned outside the soft tissue[12][13].

When this design was first developed, microgap bacteria residing between the abutment and implant were considered to be the cause of marginal bone resorption[18][19][20]. Therefore, a design was developed to export the position of this microgap outside the body[21]. However, such a design limits the customization of the patient’s emergence profile, resulting in difficulties in terms of long-term clinical results for aesthetically acceptable implant prostheses.

To maintain an adequate soft tissue seal, there must be a proper foundation of underlying bone. However, nonsubmerged-type implants can be susceptible to marginal bone resorption due to the concentration of stress at the top of the implant[22].

1.5. Materials for Abutment

Achieving soft tissue seals may also depend on the type of material used for the abutment. Abrahamsson et al. (1998) investigated the stability of soft tissue seals attached via gold alloy, dental porcelain, titanium, and aluminum oxide. In this study, marginal bone resorption occurred when seals were attached to surfaces composed of gold alloy and dental porcelain, because no soft tissue seal formed. However, when titanium and aluminum oxide were used, a soft tissue seal was achieved, and marginal bone was not absorbed[23]. Welander et al. (2008) also reported that epithelial adhesions receded around abutments attached via gold alloy[24]. Therefore, when using a UCLA abutment, the transmucosal area of the abutment is made of gold or dental porcelain, and a soft tissue seal may not be properly achieved, resulting in marginal bone resorption.

More recently, titanium and zirconia have been used as materials for abutments. One advantage of zirconia is that it is more aesthetic than titanium. While several studies show that there are no significant differences in the soft tissue seal between these two materials, zirconia has a lower fracture strength than titanium, and is thus more likely to be associated with mechanical complications[24][25][26]. In addition, adhesion of bacteria to titanium occurs less readily than to zirconia[27].

Recent advances in digital dentistry have made it possible to fabricate and restore implant-supported restorations in just a single visit. To this end, lithium disilicate and polymer-infiltrated ceramic network (PICN), which are easy to process on the day, are used as abutment material. These materials are used as implant abutments after cementing to the prefabricated titanium base with adhesive resin cement. Although little research has been conducted on the soft tissue seal of these new materials, Smallidge et al. have reported favorable epithelial cell growth on the PICN surface when PICN has a relatively low surface roughness (Ra < 0.254 µm)[28]. Another recent study reported that the surface roughness of lithium disilicate was smoother than that of PICN after the same polishing process with 6 μm diamond slurry, and the surface roughness of both of materials was unaffected by the ultrasonic scaling procedure[29]. Mehl et al. found that the adhesive resin joint connecting the titanium base and the abutment materials had neither influence on soft tissue anatomy nor on bone loss in the animal study[30]. In zirconia abutments, however, junctional epithelium was significantly shorter than titanium one-piece abutments[30]. Ongun et al. measured the mechanical properties of these hybrid-type abutments including titanium bases, showing that PICN had lower fracture resistance and weaker adhesion to resin cements compared with lithium disilicate ceramic[31]. Such mechanical failure of PICN hybrid-type abutments could also cause the soft tissue seal to break.

1.6. Detachment of Abutments

When the abutment becomes disconnected from a submerged-type implant, the soft tissue seal is broken, and microorganisms in the oral cavity can then penetrate into the tissues surrounding the implant[32]. This may result in the loss of marginal bone. Abrahamsson et al. reported that the absorption of marginal bone is doubled when the abutment is detached five times[33]. For these reasons, some clinicians have proposed the ‘one abutment-one time (OAOT)’ concept to prevent marginal bone loss[34][35]. Clinicians should be aware that during a prosthetic restoration procedure, or during the postrestoration maintenance period, it may be advantageous to minimize the process of removing the abutment.

A systematic review suggested that the OAOT protocol might preserve peri-implant bone loss and soft tissue changes for two reasons[36]. First of all, the micro-gap between abutment and implant could cause to bacterial leakage and micromotion, and this might lead to the inflammation of peri-implant soft tissue and bone resorption. The OAOT protocol could provide less micro-gap because healing or temporary abutments were installed by less preloading force (<10 Ncm) than final abutments (about 30 Ncm)[36]. Second, the OAOT protocol could reduce the disruption of the soft tissue seal around the implant abutment complex to avoid repeated the dis/reconnection of abutments[36].

1.7. Surface Modification of Abutments

Various technologies have been explored for improving the soft tissue seal at the transmucosal part of the abutment, including surface treatments, such as coating, machining, blasting, plasma spraying, etching, and laser processing[37].

Yang et al. reported that, when ultraviolet light is applied to surfaces, gingival fibroblasts proliferate more readily on surfaces made of zirconia[38]. Several studies showed that, when an abutment is laser-treated, the connective tissue is directly attached to the surface of the abutment by perpendicular fibers[39][40][41]. In addition, as the surface roughness of the transmucosal part of the abutment increases, the soft tissue seal improves. This may be because a rougher surface has a larger surface area to which the soft tissue can attach[42]. A previous study also suggested that hydrothermal treatment on titanium alloy could change minimally surface topography to enhance hydrophilicity[43]. Subsequently, the treatment contributed to the integration of epithelial cells to the surface and might facilitate the healthy epithelial tissue sealing around the transmucosal part of the implant[43].

The surface of the abutment can be modified to change the surface roughness or the free energy. While this may enhance the soft tissue seal, it can also increase plaque accumulation and the tendency for bacterial colonization to occur[44][45]. Thus, surface treatments applied to abutments should be carefully selected.

2. Hard Tissue Responses to Implant System Materials and Structures

2.1. The Bone Stimulation Theory

The soft and hard tissues surrounding the implant play complementary roles. Alveolar bone is a hard tissue that withstands the masticatory force applied to the implant, and serves to transmit it to the jawbone. The soft tissue, gingiva, and mucosa protect the alveolar bone from external irritants such as bacteria. The condition of the soft tissue is maintained by the underlying alveolar bone, and the alveolar bone is protected by the overlying soft tissue.

In general, after the restoration of an implant, between 1 and 1.5 mm of the marginal bone around the implant is absorbed during the first year owing to the application of occlusal force, and 0.2 mm is absorbed per year thereafter[46]. On the basis of the results of studies showing that marginal bone is steadily absorbed every year, it was thought that long-length implants should be beneficial for long-term predictability[46]. This was observed in long-term studies of the Brånemark system using an external connection-type implant, which is limited and inapplicable to some internal friction connections[47][48][49]. By contrast, several studies on a certain implant connection system reported that peri-implant marginal bone is increased by occlusal loading[48][49]. Because the use of long implants is likely to cause damage to important anatomical structures such as nerves, it may be more advantageous to insert short implants if warranted for long-term prognosis without marginal bone resorption.

In natural dentition, when periodontal disease occurs or a natural tooth is lost, the alveolar bone of the corresponding region is resorbed. However, when the natural tooth and the periodontal tissue are healthy, the alveolar bone is well-maintained. Except for a small amount of loss due to mechanical degeneration or aging, the alveolar bone should be well-maintained throughout life, as periodontal ligaments have been shown to transfer adequate stimulation to alveolar bone. Otherwise, disuse atrophy of the bone can occur. From this point of view, if the peri-implant bone is appropriately stimulated by the implant, the peri-implant marginal bone will not be resorbed, and the quantity and quality of the bone are likely to be preserved.

The Astra implant system (Dentsply Sirona, Charlotte, , NC, United States) involves an internal friction connection type, and was the first connection system to embody this concept. The implant–abutment connection of this system improves the amount and quality of peri-implant bone by properly transmitting the masticatory force to the surrounding alveolar bone. This hypothesis was reported by Frost[50]and can be applied to alveolar bone as well as to other bones in the human body[50]. Hence, under conditions involving the application of appropriate strain to human bone, osteoblasts are activated, which increases the amount and quality of the bone.

Before existing implant systems were widely used in clinical practice, various types and shapes of implants were developed, and most disappeared without displaying any ability to withstand mechanical or biological complications[51][52]. The Brånemark system was the first to achieve successful long-term predictable prognosis, and the consequential popularization of implants. Unlike previous plain-shaped systems, this system can deliver occlusal force to the alveolar bone because it has a thread on the implant surface[51][52]. The thread of the implant transforms the shear stress generated at the interface between the implant and the bone into compressive stress so that the appropriate stimulus can be transmitted to the bone (Figure 5). This stimulation is one of the factors that allows the bone to remain stable over the long term.

Materials 13 00072 g005 550

Figure 5. The thread function of non-threaded (left) and threaded (right) dental implants. Shear force (red arrows) is transformed into compressive force (green arrow) by the threads. Note the reduction of shear force due to the partial switch to compressive force in the magnified diagram (vertically and horizontally; not marked).

A disadvantage of these external connection-type implants is that the soft tissue seal formed around the abutment is destroyed by micromobility at the connection. The Brånemark implant represents the beginning of modern implants, and was the first system to perform well in clinical practice. However, marginal bone resorption was found to be caused by the destruction of the soft-tissue seal. To prevent such marginal bone loss, an Astra internal connection implant system was developed. This internal friction system maintains bone by applying appropriate stimulation to the bone around the implant without loss of the soft tissue seal around the implant.

2.2. Mechanism of Bone Stimulation

The connection of the Astra implant system has an internal conical shape with a slope of 11 degrees. The occlusal force applied to the abutment is transmitted to the implant through this conical connection. The masticatory force delivered to the implant thus becomes the source of the stain to stimulate the alveolar bone[53]. Thus, when the abutment receives occlusal force and sinks downward, the conical opening of the implant is opened wider, and the bone around the implant is consequentially stimulated (Figure 6). This stimulation activates osteoblasts in the alveolar bone, thereby increasing the amount and quality of alveolar bone. This increase in alveolar bone can lead to a positive change in the results of clinical procedures. This reduces the need for invasive bone grafting or the placement of excessively long implants during implant surgery. This represents a positive result for both the patient and the surgeon.

2.3. Prevention of Hard Tissue Loss

The marginal bone loss that occurs in the mobile connection of external connection-type implants does not lead directly to the failure of the implant. However, it can cause considerable complications in the tissue surrounding the implant. In general, these external connection implants are known to lose marginal bone up to the second or third thread level of the implant[54]. This bone loss can alter the properties of the overlying soft tissue, resulting in a reduction in the attached gingiva. This weakens the soft tissue seal and increases the likelihood of bacterial invasion and peri-implantitis. A detailed periodical examination of the condition of tissues around the implant is thus essential for long-term success when external connection-type implants are used.

Materials 13 00072 g006 550

Figure 6. Schematic drawing of bone stimulation. When occlusal force is applied to an internal connection-type implant, the implant around the connection expands and stimulates the surrounding bone to induce bone proliferation. (a) Internal friction connection-type implant. (b) The occlusal force is transmitted to the implant through the conical connection parts (red lines). (c) The coronal expansion of the implant (curved red arrows) becomes the source of the strain that stimulates the alveolar bone.

The issue of whether gingiva must be present for the health of natural teeth has been investigated. Clinically, even if the gingiva cannot completely wrap the tooth, it should be able to prevent the movement of the surrounding soft tissue from being transmitted to the free gingiva[55]. In other words, cheek, lip, and mandibular movements should not be transferred to the marginal gingiva. To achieve this, a certain amount of attached gingiva must be present to prevent the penetration of bacteria into the periodontal tissue[56].

This principle of natural dentition should also be borne in mind when considering implant restorations[57]. An appropriate amount of attached gingiv should be present, rather than free gingival alone, to prevent microbial invasion into tissue surrounding the implant. According to the systematic review in 2018, if there is the insufficient attaced gingiva around implant, the apically positioned flap with autogenous graft is suggested to increase the width of the gingiva[58]. However, it is not enough simply to increase the width of attached gingiva, because this surgical procedure is unable to alter the level of mucogingival junction[59]. Additional vestibuloplasty is, therefore, recommended to create "new" mucogingival junction by preventing the movement of the lips and cheeks from affecting th free gingiva.

This publication can be found here:https://www.mdpi.com/1996-1944/13/1/72

References

  1. P. Mark Bartold; Laurence J. Walsh; A. Sampath Narayanan; Molecular and cell biology of the gingiva.. Periodontology 2000 2000, 24, 28-55, 10.1034/j.1600-0757.2000.2240103.x.
  2. T. Berglundh; J. Lindhe; I. Ericsson; C. P. Marinello; B. Liljenberg; P. Thornsen; The soft tissue barrier at implants and teeth. Clinical Oral Implants Research 1991, 2, 81-90, 10.1034/j.1600-0501.1991.020206.x.
  3. I.-S. Moon; T. Berglundh; I. Abrahamsson; E. Linder; J. Lindhe; The barrier between the keratinized mucosa and the dental implant. An experimental study in the dog.. Journal of Clinical Periodontology 1999, 26, 658-663, 10.1034/j.1600-051x.1999.261005.x.
  4. Ikiru Atsuta; Yasunori Ayukawa; Akihiro Furuhashi; Takayoshi Yamaza; Yoshihiro Tsukiyama; Kiyoshi Koyano; Promotive effect of insulin-like growth factor-1 for epithelial sealing to titanium implants. Journal of Biomedical Materials Research Part A 2013, 101, 2896-2904, 10.1002/jbm.a.34608.
  5. Timothy G. Donley; William B. Gillette; Titanium Endosseous Implant-Soft Tissue Interface: A Literature Review. Journal of Periodontology 1991, 62, 153-160, 10.1902/jop.1991.62.2.153.
  6. T.R.L. Gould; L. Westbury; D.M. Brunette; Ultrastructural study of the attachment of human gingiva to titanium in vivo. The Journal of Prosthetic Dentistry 1984, 52, 418-420, 10.1016/0022-3913(84)90459-1.
  7. Hidehiro Ikeda; Masaru Shiraiwa; Takayoshi Yamaza; Masao Yoshinari; Mizuho A. Kido; Yasunori Ayukawa; Takashi Inoue; Kiyoshi Koyano; Teruo Tanaka; Difference in penetration of horseradish peroxidase tracer as a foreign substance into the peri-implant or junctional epithelium of rat gingivae.. Clinical Oral Implants Research 2002, 13, 243-251, 10.1034/j.1600-0501.2002.130303.x.
  8. Ikiru Atsuta; Yasunori Ayukawa; Ryosuke Kondo; Wakana Oshiro; Yuri Matsuura; Akihiro Furuhashi; Yoshihiro Tsukiyama; Kiyoshi Koyano; Soft tissue sealing around dental implants based on histological interpretation. Journal of Prosthodontic Research 2016, 60, 3-11, 10.1016/j.jpor.2015.07.001.
  9. Atsuta, I.; Yamaza, T.; Yoshinari, M.; Goto, T.; Kido, M.A.; Kagiya, T.; Mino, S.; Shimono, M.; Tanaka, T. Ultrastructural localization of laminin-5 (gamma2 chain) in the rat peri-implant oral mucosa around a titanium-dental implant by immuno-electron microscopy. Biomaterials 2005, 26, 6280–6287.
  10. Hidehiro Ikeda; Takayoshi Yamaza; Masao Yoshinari; Yasuyoshi Ohsaki; Yasunori Ayukawa; Mizuho A. Kido; Takashi Inoue; Masaki Shimono; Kiyoshi Koyano; Teruo Tanaka; Ultrastructural and Immunoelectron Microscopic Studies of the Peri-Implant Epithelium-Implant (Ti-6Al-4V) Interface of Rat Maxilla. Journal of Periodontology 2000, 71, 961-973, 10.1902/jop.2000.71.6.961.
  11. Zahi Badran; Ovidiu Ciobanu; Nader Hamdan; Faleh Tamimi; Mohamed-Nur Abdallah; Strategies for Optimizing the Soft Tissue Seal around Osseointegrated Implants. Advanced Healthcare Materials 2017, 6, 1700549, 10.1002/adhm.201700549.
  12. Joachim S. Hermann; Daniel Buser; Robert K. Schenk; David L. Cochran; Crestal Bone Changes Around Titanium Implants. A Histometric Evaluation of Unloaded Non-Submerged and Submerged Implants in the Canine Mandible. Journal of Periodontology 2000, 71, 1412-1424, 10.1902/jop.2000.71.9.1412.
  13. Joachim S. Hermann; John D. Schoolfield; Robert K. Schenk; Daniel Buser; David L. Cochran; Influence of the Size of the Microgap on Crestal Bone Changes Around Titanium Implants. A Histometric Evaluation of Unloaded Non-Submerged Implants in the Canine Mandible. Journal of Periodontology 2001, 72, 1372-1383, 10.1902/jop.2001.72.10.1372.
  14. Alberto Monje; Angel Insua; Hom-Lay Wang; Understanding Peri-Implantitis as a Plaque-Associated and Site-Specific Entity: On the Local Predisposing Factors.. Journal of Clinical Medicine 2019, 8, 279, 10.3390/jcm8020279.
  15. Stefano Gracis; Konstantinos Michalakis; Paolo Vigolo; Per Vult Von Steyern; Marcel Zwahlen; Irena Sailer; Internal vs. external connections for abutments/reconstructions: a systematic review. Clinical Oral Implants Research 2012, 23, 202-216, 10.1111/j.1600-0501.2012.02556.x.
  16. Yang Liu; Jiawei Wang; Influences of microgap and micromotion of implant–abutment interface on marginal bone loss around implant neck. Archives of Oral Biology 2017, 83, 153-160, 10.1016/j.archoralbio.2017.07.022.
  17. I. Abrahamsson; T. Berglundh; J. Wennstrom; J. Lindhe; The peri-implant hard and soft tissues at different implant systems. A comparative study in the dog.. Clinical Oral Implants Research 1996, 7, 212-219, 10.1034/j.1600-0501.1996.070303.x.
  18. Simonetta D'ercole; Antonio Scarano; Vittoria Perrotti; Jorge Mulatinho; Adriano Piattelli; Giovanna Iezzi; Domenico Tripodi; Implants With Internal Hexagon and Conical Implant-Abutment Connections: An In Vitro Study of the Bacterial Contamination. Journal of Oral Implantology 2014, 40, 30-34, 10.1563/aaid-joi-d-11-00121.
  19. Theofilos Koutouzis; Ruben Mesia; Nadia Calderon; Fong Wong; Shannon Wallet; The Effect of Dynamic Loading on Bacterial Colonization of the Dental Implant Fixture-Abutment Interface: An In-vitro Study.. Journal of Oral Implantology 2012, null, , 10.1563/aaid-joi-d-11-00207.1.
  20. Adriano Piattelli; Giuseppe Vrespa; Giovanna Petrone; Giovanna Iezzi; Susanna Annibali; Antonio Scarano; Role of the Microgap Between Implant and Abutment: A Retrospective Histologic Evaluation in Monkeys. Journal of Periodontology 2003, 74, 346-352, 10.1902/jop.2003.74.3.346.
  21. Daniel Buser; Regina Mericske-Stern; Karl Dula; Niklaus P. Lang; Clinical experience with one-stage, non-submerged dental implants.. Advances in Dental Research 1999, 13, 153-161, 10.1177/08959374990130010501.
  22. Stig Hansson; A conical implant-abutment interface at the level of the marginal bone improves the distribution of stresses in the supporting bone. An axisymmetric finite element analysis.. Clinical Oral Implants Research 2003, 14, 286-293, 10.1034/j.1600-0501.2003.140306.x.
  23. Lngemar Abrahamsson; Tord Berglundh; Jan Lindhe; Soft tissue response to plaque formation at different implant systems. A comparative study in the dog. Clinical Oral Implants Research 1998, 9, 73-79, 10.1034/j.1600-0501.1998.090202.x.
  24. Maria Welander; Ingemar Abrahamsson; Tord Berglundh; The mucosal barrier at implant abutments of different materials. Clinical Oral Implants Research 2008, 19, 635-641, 10.1111/j.1600-0501.2008.01543.x.
  25. Ralf J. Kohal; Dietmar Weng; Maria Bachle; Jörg R. Strub; Loaded Custom-Made Zirconia and Titanium Implants Show Similar Osseointegration: An Animal Experiment. Journal of Periodontology 2004, 75, 1262-1268, 10.1902/jop.2004.75.9.1262.
  26. Suet Yeo Soo; Nikolaos Silikas; Julian Satterthwaite; Measurement of Fracture Strength of Zirconia Dental Implant Abutments with Internal and External Connections Using Acoustic Emission.. Materials 2019, 12, 2009, 10.3390/ma12122009.
  27. Erica Dorigatti De Avila; Rafael Scaf De Molon; Bruno P. Lima; Renate Lux; Wenyuan Shi; Miguel Jafelicci Junior; Denise Madalena Palomari Spolidorio; Carlos Eduardo Vergani; Francisco De Assis Mollo Junior; Impact of Physical Chemical Characteristics of Abutment Implant Surfaces on Bacteria Adhesion. Journal of Oral Implantology 2016, 42, 153-158, 10.1563/aaid-joi-d-14-00318.
  28. Martin J. Smallidge; Jennifer V. Sabol; Cynthia Aita-Holmes; Henry Chuang; Uglas P. Dickinson; Human Gingival Epithelial Growth In Vitro on a Polymer‐Infiltrated Ceramic Network Restorative Material. Journal of Prosthodontics 2019, 28, 541-546, 10.1111/jopr.13017.
  29. Jae-Hyun Lee; Sung-Hun Kim; Jung-Suk Han; In-Sung Luke Yeo; Hyung-In Yoon; Jungwon Lee; Effects of ultrasonic scaling on the optical properties and surface characteristics of highly translucent CAD/CAM ceramic restorative materials: An in vitro study. Ceramics International 2019, 45, 14594-14601, 10.1016/j.ceramint.2019.04.177.
  30. Christian Mehl; Volker Gaβling; Stephan Schultz-Langerhans; Yahya Açil; Telse Bähr; Jörg Wiltfang; Matthias Kern; Influence of Four Different Abutment Materials and the Adhesive Joint of Two-Piece Abutments on Cervical Implant Bone and Soft Tissue. The International Journal of Oral & Maxillofacial Implants 2016, 31, 1264-1272, 10.11607/jomi.5321.
  31. Salim Ongun; Sevcan Kurtulmus-Yilmaz; Gökçe Meriç; Mutahhar Ulusoy; A Comparative Study on the Mechanical Properties of a Polymer-Infiltrated Ceramic-Network Material Used for the Fabrication of Hybrid Abutment. Materials 2018, 11, 1681, 10.3390/ma11091681.
  32. Xavier Rodriguez; Xavier Vela; Maribel Segalà; Dennis P. Tarnow; Jose L. Calvo‐Guirado; Víctor Méndez; The effect of abutment dis/reconnections on peri‐implant bone resorption: A radiologic study of platform‐switched and non‐platform‐switched implants placed in animals. Clinical Oral Implants Research 2011, 24, 305-311, 10.1111/j.1600-0501.2011.02317.x.
  33. I. Abrahamsson; T. Berglundh; J. Lindhe; The mucosal barrier following abutment dis/reconnection. An experimental study in dogs.. Journal of Clinical Periodontology 1997, 24, 568-572, 10.1111/j.1600-051x.1997.tb00230.x.
  34. Luigi Canullo; Isabella Bignozzi; Roberto Cocchetto; Maria Paola Cristalli; Giuliano Iannello; Immediate positioning of a definitive abutment versus repeated abutment replacements in post-extractive implants: 3-year follow-up of a randomised multicentre clinical trial.. European journal of oral implantology 2010, 3, 285–296, https://www.researchgate.net/profile/Isabella_Bignozzi/publication/49704601_Immediate_positioning_of_a_definitive_abutment_versus_repeated_abutment_replacements_in_post-extractive_implants_3-year_follow-up_of_a_randomised_multicentre_clinical_trial/links/0912f5058322e5d9be000000.pdf.
  35. Marco Degidi; Diego Nardi; Adriano Piattelli; One abutment at one time: non-removal of an immediate abutment and its effect on bone healing around subcrestal tapered implants. Clinical Oral Implants Research 2011, 22, 1303-1307, 10.1111/j.1600-0501.2010.02111.x.
  36. Qing-Qing Wang; Ruoxi Dai; Chris Ying Cao; Hui Fang; Min Han; Quan-Li Li; One-time versus repeated abutment connection for platform-switched implant: A systematic review and meta-analysis. PLOS ONE 2017, 12, e0186385, 10.1371/journal.pone.0186385.
  37. Paolo Ghensi; Elia Bettio; Devid Maniglio; Emiliana Bonomi; Federico Piccoli; Silvia Gross; Patrizio Caciagli; Nicola Segata; Giandomenico Nollo; Francesco Tessarolo; et al. Dental Implants with Anti-Biofilm Properties: A Pilot Study for Developing a New Sericin-Based Coating.. Materials 2019, 12, 2429, 10.3390/ma12152429.
  38. Yang Yang; Jianfeng Zhou; Xiaoqiang Liu; Miao Zheng; Jianjun Yang; Jianguo Tan; Ultraviolet light-treated zirconia with different roughness affects function of human gingival fibroblastsin vitro: The potential surface modification developed from implant to abutment. Journal of Biomedical Materials Research Part B: Applied Biomaterials 2014, 103, 116-124, 10.1002/jbm.b.33183.
  39. Gerhard Iglhaut; Vladimir Golubovic; Kathrin Becker; Henning Schliephake; Ilja Mihatovic; The impact of dis‐/reconnection of laser microgrooved and machined implant abutments on soft‐ and hard‐tissue healing. Clinical Oral Implants Research 2012, 24, 391-397, 10.1111/clr.12040.
  40. Myron Nevins; Marcelo Camelo; Marc L Nevins; Peter Schupbach; David M Kim; Connective tissue attachment to laser-microgrooved abutments: a human histologic case report.. The International Journal of Periodontics & Restorative Dentistry 2012, 32, , null.
  41. Myron Nevins; David M Kim; Sang-Ho Jun; Kevin Guze; Peter Schupbach; Marc L Nevins; Histologic evidence of a connective tissue attachment to laser microgrooved abutments: a canine study.. The International Journal of Periodontics & Restorative Dentistry 2010, 30, , null.
  42. Luigi Guida; Adriana Oliva; Maria Assunta Basile; Michele Giordano; Livia Nastri; Marco Annunziata; Human gingival fibroblast functions are stimulated by oxidized nano-structured titanium surfaces. Journal of Dentistry 2013, 41, 900-907, 10.1016/j.jdent.2013.07.009.
  43. Yasushige Sakamoto; Yasunori Ayukawa; Akihiro Furuhashi; Michimasa Kamo; Junji Ikeda; Ikiru Atsuta; Takuya Haraguchi; Kiyoshi Koyano; Effect of Hydrothermal Treatment with Distilled Water on Titanium Alloy for Epithelial Cellular Attachment.. Materials 2019, 12, 2748, 10.3390/ma12172748.
  44. Jean-Pierre Albouy; Ingemar Abrahamsson; Leif G. Persson; Tord Berglundh; Spontaneous progression of ligatured induced peri-implantitis at implants with different surface characteristics. An experimental study in dogs II: histological observations.. Clinical Oral Implants Research 2009, 20, 366-371, 10.1111/j.1600-0501.2008.01645.x.
  45. In-Sung Yeo; Ha-Young Kim; Kyung Sub Lim; Jung-Suk Han; Implant Surface Factors and Bacterial Adhesion: A Review of the Literature. The International Journal of Artificial Organs 2012, 35, 762-772, 10.5301/ijao.5000154.
  46. T Albrektsson; G Zarb; P Worthington; A R Eriksson; The long-term efficacy of currently used dental implants: a review and proposed criteria of success.. The International Journal of Oral & Maxillofacial Implants 1986, 1, 11–25, https://www.researchgate.net/profile/George_Zarb/publication/19633876_The_long-term_efficacy_of_currently_used_dental_implants_A_review_and_proposed_criteria_of_success/links/56570d5008aeafc2aac0b605/The-long-term-efficacy-of-currently-used-dental-implants-A-review-and-proposed-criteria-of-success.pdf.
  47. Rodrigo Antonio De Medeiros; Eduardo Piza Pellizzer; Aljomar José Vechiato Filho; Daniela Micheline Dos Santos; Emily Vivianne Freitas Da Silva; Marcelo Coelho Goiato; Evaluation of marginal bone loss of dental implants with internal or external connections and its association with other variables: A systematic review. The Journal of Prosthetic Dentistry 2016, 116, 501-506, 10.1016/j.prosdent.2016.03.027.
  48. Richard M. Palmer; Paul J. Palmer; Brian J. Smith; A 5-year prospective study of Astra single tooth implants.. Clinical Oral Implants Research 2000, 11, 179-182, 10.1111/j.1600-0501.2000.tb00012.x.
  49. L Puchades-Roman; R M Palmer; P J Palmer; L C Howe; M Ide; R F Wilson; A clinical, radiographic, and microbiologic comparison of Astra Tech and Brånemark single tooth implants.. Clinical Implant Dentistry and Related Research 2000, 2, , null.
  50. Harold M. Frost; Perspectives: bone's mechanical usage windows. Bone and Mineral 1992, 19, 257-271, 10.1016/0169-6009(92)90875-e.
  51. Daniel Buser; Lars Sennerby; Hugo De Bruyn; Modern implant dentistry based on osseointegration: 50 years of progress, current trends and open questions. Periodontology 2000 2016, 73, 7-21, 10.1111/prd.12185.
  52. Thomas D. Taylor; Twenty years of progress in implant prosthodontics. The Journal of Prosthetic Dentistry 2002, 88, 89-95, 10.1016/s0022-3913(02)00051-3.
  53. Luis Gustavo Oliveira De Vasconcellos; Alberto Noriyuki Kojima; Renato Sussumu Nishioka; Luana Marotta Reis De Vasconcellos; Ivan Balducci; Axial Loads on Implant-Supported Partial Fixed Prostheses for External and Internal Hex Connections and Machined and Plastic Copings: Strain Gauge Analysis. Journal of Oral Implantology 2015, 41, 149-154, 10.1563/aaid-joi-d-10-00174.
  54. D.P. Tarnow; S.C. Cho; S.S. Wallace; The Effect of Inter-Implant Distance on the Height of Inter-Implant Bone Crest. Journal of Periodontology 2000, 71, 546-549, 10.1902/jop.2000.71.4.546.
  55. Marlize T. Friedman; Pauline M. Barber; Nicola J. Mordan; Hubert N. Newman; The “Plaque-Free Zone” in Health and Disease: A Scanning Electron Microscope Study. Journal of Periodontology 1992, 63, 890-896, 10.1902/jop.1992.63.11.890.
  56. Niklaus P. Lang; Harald Löe; The Relationship Between the Width of Keratinized Gingiva and Gingival Health. Journal of Periodontology 1972, 43, 623-627, 10.1902/jop.1972.43.10.623.
  57. Daniel S Thoma; Sven Mühlemann; Ronald E Jung; Critical soft-tissue dimensions with dental implants and treatment concepts.. Periodontology 2000 2014, 66, 106-118, 10.1111/prd.12045.
  58. William V. Giannobile; Ronald E. Jung; Frank Schwarz; The Groups Of The 2Nd Osteology Foundation Consensus Meeting; Evidence-based knowledge on the aesthetics and maintenance of peri-implant soft tissues: Osteology Foundation Consensus Report Part 1-Effects of soft tissue augmentation procedures on the maintenance of peri-implant soft tissue health. Clinical Oral Implants Research 2018, 29, 7-10, 10.1111/clr.13110.
  59. Anja Anjcamo; Axel Bergenholtz; Anders Hugoson; Jukka Ainamo; Location of the mucogingival junction 18 years after apically repositioned flap surgery. Journal of Clinical Periodontology 1992, 19, 49-52, 10.1111/j.1600-051x.1992.tb01148.x.
More
Information
Contributors MDPI registered users' name will be linked to their SciProfiles pages. To register with us, please refer to https://encyclopedia.pub/register : , , , ,
View Times: 2.4K
Revisions: 5 times (View History)
Update Date: 27 Oct 2020
1000/1000